10
Maintenance
10.1
Safety
Requirements for the maintenance personnel
•
Only allow an authorized electrician to work on the electrical system or electrical
subassemblies.
•
The technician must be familiar with the exhaust gas technology of the operator's
plant (hazard by overpressure and toxic and hot flue gases) and be able to avoid
hazards when working on gas ducts.
•
The technician must be familiar with handling compressed gas cylinders (test
gases).
•
The technician must be able to avoid hazards caused by noxious test gases.
•
The technician must be familiar with gas lines and their screw fittings (be able to
ensure gas-tight connections).
10.2
Maritime Hotline
If you have questions about malfunctions or maintenance work, you can contact the
Maritime Hotline (24/7).
Phone: +49 7681 2024194
E-mail: [email protected]
10.3
Data backup and data recovery
The SOPAS data backup must be carried out before and after each commissioning,
maintenance and repair for a complete documentation of the analyzer.
By importing the SOPAS data backup, an error can be reset during execution.
The data backup can also be opened off-line. This allows an appropriate expert to
identify pending errors in a current backup and define the next steps to fix them.
10.3.1
Backing up the data
Overview
A data backup must be carried out before as well as after each commissioning, mainte‐
nance and repair work.
A data backup must be performed for all modules of the measuring device.
Prerequisites
•
SOPAS ET is connected to the measuring device.
•
Logged in with the Authorized Operator user level.
Procedure
1.
Click on the three dots of the corresponding module and select “Export”.
2.
Select the appropriate destination folder and click “Save”.
10.3.2
Recovering the data
Overview
Data recovery can be used to correct erroneous data or update the system.
MAINTENANCE
10
8021339/1CHC/V1-1/2021-07 | SICK
O P E R A T I N G I N S T R U C T I O N S | MARSIC280
69
Subject to change without notice