7.3
Changing the IP address on the BCU
Overview
Changing the IP address of the computer is preferable. If this is not possible, the IP
address can be changed via the BCU.
If the IP address of the device is changed on the BCU, it must be restored at the end of
the work.
Procedure
1.
Open the menu.
2.
Select “Login”.
3.
Select “Authorized Operator”.
4.
Enter password and click “Save”.
5.
Press “Back”.
6.
Press “Parameter”.
7.
Press “Communication”.
8.
Select “LAN”.
9.
Note the displayed IP address during maintenance and repair work.
10. Press “Set”.
11. Change IP address to the number range of the computer.
12. Press “Save”.
✓
The IP address is changed.
7.4
Adjusting
7.4.1
Setting the test gas
Overview
Before adjustment, the exact test gas concentrations must be entered in the Test Gas
Table.
Procedure
1.
Check the cylinder concentration on the test gas cylinder.
2.
Press “Menu”.
3.
Select menu “3 Adjustments”.
4.
Select menu “5 Settings”.
5.
Select menu “1 Test gases”.
6.
Select the appropriate component for which a test gas setting is to be made.
7.
Select menu “5 Usage”.
8.
Select the appropriate component.
9.
Select menu “3 Concentration”.
10. Enter the test gas concentration using the arrow keys.
11. Press “Save“ to confirm the input.
7.4.2
Setting the test gas with SOPAS ET
Overview
Before adjustment, the exact test gas concentrations must be entered in the Test Gas
Table.
Procedure
1.
Check the cylinder concentration on the test gas cylinder.
2.
Open the BCU module with a double-click.
3.
Open the “Test gas table” menu under “Parameter”.
7
COMMISSIONING
50
O P E R A T I N G I N S T R U C T I O N S | MARSIC280
8021339/1CHC/V1-1/2021-07 | SICK