TROUBLESHOOTING
11
Troubleshooting
11.1
Error message on the display
Error
Cause
Possible solution
!InvEc &
level present
AutCal not executed; interference superimposes
medium reflection.
Perform commissioning
(See
“7.1 Quick commissioning (with factory
TrsHld setting is not suitable for the medium.
Perform advanced commissioning
(See
).
!InvEc &
empty tank
Probe length configured incorrectly.
Check probe length and compare against
configuration in EXPRT-Config-LENGTH.
Probe not available.
Check the probe.
!ATTNT
A parameter was written outside of the valid value
range and therefore adjusted
Rewrite the value in the valid range.
Another parameter was automatically adjusted
due to a dependency (SPx, RPx).
Check the parameter again.
!WRONG
Incorrect password entered.
Enter the correct password.
!NoCal
Information: The AutCal process and/or foam
calibration was rejected because the probe length,
teach-in depth or measuring mode was changed.
Perform commissioning again if necessary.
!Denid
AutCal was called up in Foam sensor mode.
AutCal is available only in Pulse mode. Perform
foam calibration in Foam mode.
CalPIs
No valid calibrations for CalEmp and CalMed.
Perform foam calibration.
!CalOk
The teach-in process was successful.
!NoSig
AutCal failed.
Repeat commissioning.
!faild
Foam-CalEmp or FoamCalMed menu item failed.
Follow the foam commissioning instructions.
!SC-Q1
!SC-Q2
!SC-Q3
!SC-Q4
!SC-Qa
Short-circuit at the output.
Remove short-circuit.
Load resistance at the output is too low.
Increase the load resistance.
!IOLOf
Supply voltage too low for IO-Link communication. Increase the supply voltage to achieve the desired
functionality.
!QaOff
Supply voltage too low for analog output.
Increase the supply voltage to achieve the desired
functionality.
!QxOff
Supply voltage too low for switching outputs.
Increase the supply voltage to achieve the desired
functionality.
!QaOvf
The ohmic load at the analog current output Qa is
too high.
Reduce the load at Qa.
The analog current output Qa is not wired.
Connect the load to Qa.
!MaskZ
Interference/pulse exceeds value for MaskTr.
Increase MaskTr or identify and eliminate
interference.
!Range
The maximum allowable measuring range was
exceeded. Measurement in this configuration is
not possible.
Reduce the probe length and/or coaxial cable
length; see
“8.4.10 Changing the coaxial cable
.
!Cable
The coaxial cable is damaged/defective.
Replace the coaxial cable.
The coaxial cable length was configured incorrectly.
See
“8.4.10 Changing the coaxial cable length”.
The display only
shows RUN. It is
otherwise empty.
The DspVal menu parameter is at QaBarG and the
level is below QALOW.
Change QALOW or DspVal.