FLOWSIC600 · Technical Information · 8010125 V 4.0 · © SICK AG
Appendix
Su
bj
ec
t t
o c
han
ge
w
ith
ou
t n
ot
ic
e
AO Value at 4mA
0 m3/h
Volumetric flow rate at minimum current
AO at error
22 mA
Current output in "Error" state
AO damping
0 s
Averaging period for the current output in order to smooth
peaks
Settings
SerialInterface
Submenu "Serial Interface" with the parameters for the serial
interface.
Serial Interface
Reset RS485/1?
Resetting the parameter serial interface 1 to the default val-
ues. Defaults are: Baudrate 9600, protocol ASCII (MODBUS
ASCII), 8, n, 1. Master burst mode off.
Serial Interface
Reset Service?
Resetting the parameter serial interface 2 to the default values
(see above).
Serial Interface
Reset RS485/2?
Resetting the parameter serial interface 2 to the default values
(see above).
Settings
LCD Display
Submenu "LCD Display" with language parameter for the LCD
display.
Language
English
Set parameter "Language" to
- English
- German
- French
- Russian
Main Menu
Submenu
Submenu
Notes