FLOWSIC100 Flare-XT · Operating Instructions · 8023761/V1-0/2020-10 · © SICK Engineering GmbH
Installation FLSE100-XT
Check the following points before installation
● Sender/receiver units to be fitted for a particular measuring point must belong to the
same system to ensure the maximum measurement accuracy possible. Exchanging
structurally identical sender/receiver units from different measuring systems is not
allowed.
● Sender/receiver unit pairs are matched and can only be exchanged in pairs for identical
units.
● Sender/receiver units from one system are marked with sequential serial numbers
(printed on the device label).
The FLSE100-XT Master always has the lower number and the FLSE100-XT Slave the
higher number.
● The flange connections of the sender/receiver units and nozzles must be compatible.
● The flange connections of the nozzles must be free of welding beads on the inside.
NOTICE:
The deformation characteristic of the flange gasket has an influence on the
geometry of the installation and therefore on the uncertainty of the
measurement. SICK recommends:
● Only use the same gasket type as the original delivery.
● Apply a tightening torque according to the installed gasket,