64
DUSTHUNTER T · Operating Instructions · 8012428 V2.2 · © SICK AG
Start-up and Parameter Settings
Su
bj
ec
t t
o c
han
ge
w
ith
ou
t n
ot
ic
e
– Onsite installation of a "zero tube" (tube with suitable diameter with flanges)
Figure 51
Creating a smoke-free path with zero tube (shown for DUSTHUNTER T100)
Using the associated connection cables, connect the sender/receiver unit to the MCU
and also the reflector to the sender/receiver unit for DUSTHUNTER T200.
Connect the MCU to the supply voltage.
Start the SOPAS ET program and connect to the measuring system (
).
Select the appropriate device file for the installed device (DH T50, DH T100, DH T200)
and move it to the "Project tree" window (
).
Set the sender/receiver unit to "Maintenance" mode and enter the Level 1 password (
Figure 52
Setting "Maintenance" mode
Clean the optical surfaces on sender/receiver unit and reflector (
).
Wait approx. 30 min before starting the following work (measuring system must have
reached operating temperature).
A
Ø76
Ø76
The respective device type connected is displayed automatically