74
DUSTHUNTER T · Operating Instructions · 8012428 V2.2 · © SICK AG
Start-up and Parameter Settings
Su
bj
ec
t t
o c
han
ge
w
ith
ou
t n
ot
ic
e
4.4
Setting standard parameters
4.4.1
Assigning the MCU to the sender/receiver unit
The MCU must be set to the sender/receiver unit to be connected. A malfunction is
reported in case of a mismatch. Assignment must be made after installation when the
setting is not possible at the factory (e.g., when several devices are delivered at the same
time or when the MCU is swapped later). The following steps are then necessary:
Connect the measuring system to the SOPAS ET program, select device file "MCU" and
move it to the "Project tree" window (
).
Set the Level 1 password (
) and set the measuring system to "Mainte-
nance" mode (activate the "Maintenance on/off" checkbox in the "Maintenance / Main-
tenance / Operation" directory and click "Set State").
Figure 62
Setting "Maintenance" mode
Change to the "Configuration / Application selection" directory (
).
The basic type of the sender/receiver unit connected is displayed in the "Connected
variant" window (field "Application selection"). Click "Save selection" to assign to the
MCU
.
The sender/receiver unit must be connected to the MCU.