Assembly and Installation
DUSTHUNTER T · Operating Instructions · 8012428 V2.2 · © SICK AG
39
Su
bj
ec
t t
o c
han
ge
w
ith
ou
t n
ot
ic
e
Using a suitable cable (
), the MCU-N control unit (without integrated purge air
supply) can be located up to 1000 m away from the sender/receiver unit. We therefore
recommend fitting the MCU in a control room (measuring station or similar) to ensure free
access to the MCU. This considerably simplifies communication with the measuring system
in order to set parameters or to locate malfunction or error causes.
It is advantageous to provide weather protection (tin roof or similar), to be made onsite, for
use outdoors.
Requirements when using the MCU-P control unit
The following is applicable in addition to the general specifications:
●
Install the MCU-P control unit at a location with clean air whenever possible. The air
intake temperature must correspond to specifications in the Technical data (
). In unfavorable conditions, lay an air intake hose to a location with better condi-
tions.
●
The purge air hoses to the sender/receiver unit and reflector should be as short as pos-
sible.
●
Whenever possible, lay the purge air hoses so that no water can collect.
●
We recommend using the optional external purge air unit when the sender/receiver unit
and reflector are more than 10 m away from the control unit.