108
DUSTHUNTER T · Operating Instructions · 8012428 V2.2 · © SICK AG
Maintenance
Su
bj
ec
t t
o c
han
ge
w
ith
ou
t n
ot
ic
e
5.2.2
Maintenance on the reflector
DUSTHUNTER T50
Set the measuring system to "Maintenance" mode (
Loosen the CAMLOCK clips of reflector (1) and remove reflector (2).
Press purge air pipe (3) with flat screwdriver out of housing (4) and remove.
Figure 89
Reflector DHT-R5x
Carefully clean glass pane with an optics cloth.
Push the purge air pipe into the enclosure again and ensure centrical alignment.
Put on the reflector and fasten with CAMLOCK clips
Resume Measuring mode.
DUSTHUNTER T100
Set the measuring system to "Maintenance" mode.
Loosen the knurled screws and swivel the housing to the side.
Close the mounting flange with cover (
Carefully clean glass pane with an optics cloth.
Remove the cover from the mounting flange, swivel the housing back and lock with the
knurled screws.
Resume Measuring mode.
1
1
2
4
3