MAINTENANCE
6
6.2
Maintenance on the sender/receiver unit
Clean the outside of the sender/receiver unit in regular intervals. Remove deposits with
water or mechanically using suitable auxiliary means.
Work to be performed
▸
Loosen CAMLOCK clips (1) of the sender/receiver unit and swivel electronics unit (2) to
the side.
▸
Check mounting flange (5) and purge air connections (
see “Sender/receiver unit DHSB-
) for contamination and clean when necessary.
▸
Carefully clean sender optics (3) and receiver optics (4) with an optics cloth or cotton
swabs.
▸
Reassemble the sender/receiver unit again.
▸
Resume Measuring mode.
Fig. 30: Cleaning the optical interfaces
NOTE:
▸
Do not damage any device parts during maintenance work.
▸
Do not interrupt the purge air supply.
WARNING: Danger from exhaust gas
The sender/receiver unit must be opened for cleaning. It is possible that gas escapes in
plants with overpressure in the duct.
▸
Take suitable protection measures or open the sender/receiver unit only
when the plant is at a standstill.
CAMLOCK clips
Receiver optics
Electronics unit
Mounting flange
Sender optics