Operating Instructions 

Chapter 

7

 

 

DME5000 

8009813/ZJV1/2017-05 

© SICK AG • Germany • Subject to change without notice 

91

 

Menu Structure 

 

Profile 

Encoder 

This profile corresponds to the standard encoder profile, 
for description, see section 13.2 "PROFIBUS Interface"

SICK (default) 

This profile is based on the standard encoder profile. In 
addition to the measured value, warning, condition and 
error information are transmitted. 
For description, see section 13.2 "PROFIBUS Interface"

Bus Address 

Addr. 006 (default) 

Set PROFIBUS address 
(001 … 125) 

 
Commissioning example with Siemens Step 7 see chapter 13 "Annex"
 

Hiperface 
BusAddress 

Add 00 (default) 

Setting of the Hiperface address (area 0 … 31) 

Baud rate 

0.6 kBd 

Baud rate setting 

1.2 kBd   

 

2.4 kBd   

 

4.8 kBd   

 

9.6 kBd (default)   

 

19.2 kBd   

 

38.4 kBd 

 

Data 

8, o, 1 

8-data bit, odd parity, 1 stop bit 

8, e, 1 (default) 

8-data bit, even parity, 1 stop bit 

8, o, 1 

8-data bit, no parity, 1 stop bit 

Timeout 

X1 

1*11/baud rate 

X4 (default) 

4*11/baud rate 

Monitor 

On 

Serial monitor interface in operation, multifunctional 
outputs MF1 and MF2 not in operation. Enables 
parameterization of the DME5000 with the interface 
adapter (see section 13.4 "Interface Adapter") or 
monitoring of measured values via MF1 and MF2. 

Off (default) 

Multifunctional outputs MF1 and MF2 in operation as set 
in item 3.3/3.4. 

 

 
 

Note 

 
 

 

 

 

 

Summary of Contents for DME5000

Page 1: ...DME5000 Entfernungs Messger t Distance Measuring Device BETRIEBSANLEITUNG OPERATING INS TRUCTIONS D GB...

Page 2: ...ndeten Rechte bleiben bei der Firma SICK AG Eine Vervielf ltigung des Werkes oder von Teilen dieses Werkes ist nur in den Grenzen der gesetzlichen Bestimmun gen des Urheberrechtsgesetzes zul ssig Eine...

Page 3: ...5 2 2 PROFIBUS Terminierung 15 6 Bedienung 16 6 1 Aufbau des Bedienfeldes 16 6 2 Bedienung 16 6 3 Bedienfeld 16 7 Men struktur 19 7 1 Flussdiagramm 19 7 2 Erkl rungen zur Men struktur 19 8 Beispiel 2...

Page 4: ...Schnittstelle 59 13 6 1 Protokoll 60 13 6 2 Befehle 60 13 6 3 Beispiele f r Befehle Standard Protokoll 61 13 7 DeviceNet Schnittstelle 62 13 7 1 Allgemeines 62 13 7 2 Konfiguration 62 13 7 3 Datenaus...

Page 5: ...tern Vorteile bestimmter Einstellungen und helfen Ihnen den optimalen Nutzen aus dem DME5000 zu ziehen Warnhinweise Lesen und befolgen Sie diese sorgf ltig Warnhinweise sollen Sie vor Gefahren sch tze...

Page 6: ...halten 8009813 ZJV1 2017 05 Bestimmungsgem e Verwendung 2 Bestimmungsgem e Verwendung Das DME5000 ist ein optoelektronischer Sensor der zum Erfassen von Distanzen zu einer linear bewegten Reflexionsfo...

Page 7: ...achten Sie die nationalen Sicherheits und Unfallverh tungsvorschriften Reparaturen d rfen nur vom Hersteller durchgef hrt werden Eingriffe und nderungen am Ger t sind unzul ssig Verdrahtungsarbeiten f...

Page 8: ...Laserstrahls Der Empf nger des DME empf ngt das vom Reflektor zur ckgespiegelte Licht und ermit telt die Distanz zwischen Sensor und Reflektor mittels Lichtlaufzeitmessung Die gemessene Distanz wird...

Page 9: ...alterung erm glicht eine einfache Justierung in X und Y Richtung Die Reflektorgr e ist so zu w hlen dass der Lichtfleck bei Vibrationen nicht vom Reflektor abwandert Ist der Reflektor am bewegten Teil...

Page 10: ...egebenenfalls den Lichtfleck in die Mitte des Reflektors nach Die im Display Infofeld 1 angezeigte Signald mpfung gibt Aufschluss ber das aktuelle Empfangssignal Bei sauberen Grenzfl chen sollten die...

Page 11: ...dnung Entfernungs Messger t mit benachbarter Daten bertragung F r die Daten bertragungs Lichtschranken der Baureihe ISD300 ISD400 1xxx ISD400 6xxx ist unabh ngig von der maximalen Reichweite smax ein...

Page 12: ...009813 ZJV1 2017 05 Inbetriebnahme 5 2 Elektrischer Anschluss DME5000 nach Anschlussschema anschlie en Verdrahtungshinweise beachten siehe Abschnitt 5 2 1 Stecker und Kabel siehe Abschnitt 9 2 2 Abb 6...

Page 13: ...5000 8009813 ZJV1 2017 05 SICK AG Deutschland Irrt mer und nderungen vorbehalten 13 Inbetriebnahme Abb 8 Anschlussschema DME5000 xx4 DeviceNet Abb 9 Anschlussschema DME5000 xx5 DeviceNet Abb 10 Anschl...

Page 14: ...nden werden Potenzialausgleichstr me ber den Kabelschirm sind durch geeignete Ma nahmen zu vermeiden Eine st rsichere Daten bertragung ist nur bei Verwendung von abgeschirmten Leitungen mit paarweise...

Page 15: ...mer und nderungen vorbehalten 15 Inbetriebnahme 5 2 2 PROFIBUS Terminierung Abb 14 Keine Schweine schw nzchen Schirm kurz und vollfl chig anbinden BEIDE Seiten erden Abb 15 Schirmanschluss bei Kunstst...

Page 16: ...zw Funktionen 6 2 Bedienung Das DME5000 ist mit einer Folientastatur ausgestattet Bedienen Sie den Eingabebereich nur mit den Fingern oder mit einem geeigneten Zeigeger t Benutzen Sie keine spitzen od...

Page 17: ...lmeldung Empfangssignalpegel im Grenzbereich Empfangspegel OK LSR Laser Vorausfallmeldung Messlaser Laser OK PLB Plausibilit t Messfehler Ursache z B Licht strahlunterbrechung Verschmutzung Optische e...

Page 18: ...er Parameter ist ung ltig vorher g ltiger Parameter bleibt gespeichert Ursache z B unteres Schaltlimit oberes Schaltlimit Abhilfe Parameter korrigieren Infofeld 2 Beispiel 010 000 m Aktueller Messwert...

Page 19: ...s sein Es darf keine Buskommunikation aktiv sein nderungen der Busadresse sind erst nach einem Ger teneustart aktiv Wird die Eingabe des Code bersprungen oder eine andere Zahl als 314 eingegeben werde...

Page 20: ...n On default SSI Schnittstelle in Betrieb RS 422 Schnittstelle nicht in Betrieb Off RS 422 Schnittstelle in Betrieb SSI Schnittstelle nicht in Betrieb Die RS 422 Schnittstelle erm glicht das Para metr...

Page 21: ...mens Step 7 siehe Kapitel 13 Anhang Hiperface BusAddress Add 00 default Einstellung der Hiperface Adresse Bereich 0 31 Baudrate 0 6 kBd Einstellung der Baudrate 1 2 kBd 2 4 kBd 4 8 kBd 9 6 kBd default...

Page 22: ...d STX 8122 sign 7 0 9 ETX CRLF sign 7 0 9 CR LF CP0 sign 7 0 9 CP1 Nicht belegt F r eine detaillierte Beschreibung siehe Abschnitt 13 6 RS 422 Schnittstelle DeviceNet Baudrate 125 default Einstellung...

Page 23: ...et siehe 3 3 3 Service MF1 wird als Serviceausgang verwendet siehe 3 3 4 Sleepmodus MF1 wird als Eingang f r die Aktivierung des Sleepmodus Laser aus verwendet Quittierung ber Ready Status Funktionsbe...

Page 24: ...s eingestelltes Temperaturlimit Ger teinnentemperatur oberes eingestelltes Temperaturlimit Off LowerTemp On default Untertemperatur berwachung Ger teinnentemperatur unteres eingestelltes Temperaturlim...

Page 25: ...abnehmende Distanz MF2 Parametrierbarer Ausgang Active State Parametrierung der Schaltlogik Active 1 HIGH Pegel bei aktivem Ausgang Schlie er NO Active 0 LOW Pegel bei aktivem Ausgang ffner NC Functio...

Page 26: ...ingestelltes Temperaturlimit Ger teinnentemperatur oberes eingestelltes Temperaturlimit Off LowerTemp On default Untertemperatur berwachung Ger teinnentemperatur unteres eingestelltes Temperaturlimit...

Page 27: ...ter 3 1 6 Period Length Einstellbar 50 5000 m default 100 m bzw 125 m bei Hiperface Variante Offset Der Offsetwert wird zum intern ermittelten DME Messwert addiert und wirkt auf alle Ausg nge und das...

Page 28: ...0 28 SICK AG Deutschland Irrt mer und nderungen vorbehalten 8009813 ZJV1 2017 05 Men struktur Fehlerausgabe mehr m glich ist Temperatur Anzeige der aktuellen Ger teinnentemperatur Reset NO YES Kein Re...

Page 29: ...t und Weise einstellen W hrend der Parameter nderung muss der Reflektor im Sichtfeld des Sensors sein Schritt 1 Legen Sie die Betriebsspannung an Anzeige Schritt 2 Rufen Sie mit das Men CODE auf Anzei...

Page 30: ...itt 5 Gehen Sie mit eine Stelle weiter und geben Sie mit die gew nschte 2 Ziffer z B 1 ein Anzeige Schritt 6 Gehen Sie mit eine Stelle weiter und geben Sie mit die gew nschte 3 Ziffer z B 4 ein Anzeig...

Page 31: ...eichnung DME5000 xx1 DME5000 xx3 1 x M16 Stecker SSI RS 422 DME5000 xx2 1 x M16 2 x M12 Stecker PROFIBUS DME5000 xx4 DME5000 xx7 1 x M16 1 x M12 Stecker DeviceNet 2 x M12 Stecker Hiperface DME5000 xx5...

Page 32: ...17 05 Technische Daten 9 2 Zubeh r 9 2 1 Reflektoren Reflektor 0 3 x 0 3 m 2 Diamond Grade montiert 1 Reflektor 0 6 x 0 6 m 2 Diamond Grade montiert 1 Typ PL240DG Bestell Nr 1017910 Typ PL560DG Bestel...

Page 33: ...L240DG H Bestell Nr 1022926 Typ PL560DG H Bestell Nr 1023888 Reflexionsfolie Diamond Grade Gr e konfektionierbar Reflexionsfolie Diamond Grade Bogen 749 x 914 mm 2 Typ REF DG Bestell Nr 4019634 Typ RE...

Page 34: ...gsdose M16 8 polig gewinkelt 5 m geschirmt Leitungsdose M16 8 polig gewinkelt 10 m geschirmt Typ DOL 1608 W05MA Bestell Nr 2026743 Typ DOL 1608 W10MA Bestell Nr 2027194 Leitung 2 x 0 5 mm 2 6 x 0 25 m...

Page 35: ...S Leitung 2 x 0 34 mm 2 Meterware Typ PR STE 1205 G Bestell Nr 6021354 Typ LTG 2102 MW Bestell Nr 6021355 Temperaturbereich bewegt 5 80 C Festverlegt 40 80 C Mantel PUR violett 8 mm Schirm AL PT Folie...

Page 36: ...GA Bestell Nr 6027533 Leitungsdose M12 5 polig gerade Leitung 6 m Dropcable A cod Verbindungsleitung 6 m Dose M12 5 polig gerade Stecker M12 5 polig gerade Dropcable A cod Schirm auf Pin 1 Typ DOL 12...

Page 37: ...r 7 8 5 polig gerade PG16 konfektionierbar Typ DOS 7805 GK Bestell Nr 6028331 Typ STE 7805 GK Bestell Nr 6028332 Leitung 2 x 0 34 mm 2 V V bzw 2 x 0 25 mm 2 CANH CANL Adern paarweise verdrillt und mit...

Page 38: ...eitungsdose M12 5 polig PUR gerade Schirm auf R ndelmutter A cod 2 m DOL 1208 G02MAH1 Bestell Nr 6032448 5 m DOL 1205 G05MAC Bestell Nr 6036384 5 m DOL 1208 G05MAH1 Bestell Nr 6032449 10 m DOL 1205 G1...

Page 39: ...ngen vorbehalten 39 Technische Daten 9 2 3 Befestigungstechnik Ausrichthalterung DME5000 Edelstahl 1 4571 Typ BEF AH DME5 Bestell Nr 2027721 9 2 4 Wetterschutzgeh use Wetterschutzgeh use DME5000 ALMG3...

Page 40: ...Deutschland Irrt mer und nderungen vorbehalten 8009813 ZJV1 2017 05 Technische Daten 9 2 5 Umlenkspiegel Umlenkung Lichtaustritt nach oben Typ USP DME5 Bestell Nr 2027710 9 2 6 Schnittstellenadapter S...

Page 41: ...h use f r DME5000 Technische Daten Bestell Nr 6036180 Geh usematerial Glasfaserverst rkter Kunststoff GFK Betriebsumgebungs temperatur 20 75 C kurzzeitig 80 C Versorgungsspannung DC 24 V 20 Stromaufna...

Page 42: ...aserklasse 2 EN 60825 1 C D R H Lebensdauer bei 25 C MTTF 50 000 h Lichtfleckdurchmesser Entfernung max 130 mm 70 m Schaltausg nge MF1 MF2 B push pull Output MF1 MF2 HIGH UV 3 V LOW 2 V Input MF1 4 HI...

Page 43: ...60825 1 C D R H Lebensdauer bei 25 C MTTF 50 000 h Lichtfleckdurchmesser Entfernung max 270 mm 150 m Schaltausg nge MF1 MF2 B push pull Output MF1 MF2 HIGH UV 3 V LOW 2 V Input MF1 4 HIGH 12 V LOW 3 V...

Page 44: ...60825 1 C D R H Lebensdauer bei 25 C MTTF 50 000 h Lichtfleckdurchmesser Entfernung max 550 mm 300 m Schaltausg nge MF1 MF2 B push pull Output MF1 MF2 HIGH UV 3 V LOW 2 V Input MF1 4 HIGH 12 V LOW 3...

Page 45: ...3 ZJV1 2017 05 SICK AG Deutschland Irrt mer und nderungen vorbehalten 45 Wartung 10 Wartung Das DME5000 ist wartungsfrei Wir empfehlen in regelm igen Abst nden Linse und Reflektor zu reinigen Verschra...

Page 46: ...bereithalten ATT Anzeige 32H Vorausfallmeldung Verschmutzung Die Messwerte sind noch OK optische Grenzfl chen reinigen Reflektor Objektiv TMP Anzeige 1E Ger teinnentemperatur im Grenzbereich Umgebungs...

Page 47: ...Innentemperatur 80 C Ger t abk hlen Auswirkung Messwertausgabe wird auf 0 gesetzt Problem Ursache Ma nahme Parameter nderungen ber Bedientasten werden nicht dauerhaft gespeichert W hrend der nderung w...

Page 48: ...aximale Verfahrgeschwindig keit berpr fen Lichtweg zum Reflektor unterbrochen Lichtfleck bei Verfahren immer auf dem Reflektor zu sehen Auswirkung Messwertausgabe wird auf 0 gesetzt PLB Anzeige leucht...

Page 49: ...eingesetzt werden um die einzelnen Bussegmente zu verbinden Die max Leitungsl nge ist abh ngig von der bertragungsgeschwindigkeit siehe Tabelle 2 Die angegebene Leitungsl nge kann durch den Einsatz v...

Page 50: ...wischen 70 kHz und 500 kHz und ist abh ngig von der Leitungsl nge Leitungsl nge m bertragungsrate kBaud 25 500 50 400 100 300 200 200 400 100 Impulsdiagramm der Daten bertragung m gespeicherte paralle...

Page 51: ...e und ist die Standard Schnittstelle f r Motorfeedback Systeme von SICK STEGMANN Diese Schnittstelle wurde speziell auf die Anforderungen der digitalen Antriebsregelung entwickelt und bietet dem Anwen...

Page 52: ...Bei Variante DME5000 xx3 kann der Preset alternativ hierzu ber ein RS 422 Befehl erfolgen 2 Statischer Preset Ausl sung ber Multifunktionseingang MF1 Variante DME5000 xx2 bei Verwendung des GSD Modul...

Page 53: ...alisierungsposition Das Ausl sen des Presets kann je nach Ger tevariante folgenderma en erfolgen DME5000 xx1 xx4 xx7 MF1 DME5000 xx3 MF1 bzw RS 422 Kommando DME5000 xx2 MF1 bzw PROFIBUS Kommando Funkt...

Page 54: ...eingegeben wurden Ausnahme PROFIBUS Adresse und Offset bei parametriertem Preset Mode SICK Class 1 siehe ENCODER Class 1 Class 2 4 Byte Input 4 Byte Output Daten ger tespezifische Parameter sind ber...

Page 55: ...Einheit entsprechend Aufl sung Wertebereich 250 m 13 2 3 Diagnosedaten bei aktiviertem Parameter Extended Diagnostics DDLM_Slave_Diag Encoder Profil Diagnose Class 1 Octet 1 16 Octet 7 16 Class 2 Octe...

Page 56: ...rausfallmeldung Bit 1 7 Nicht verwendet Octet 21 Warnings Bit 1 Ger teinnentemperatur Bit 2 Verschmutzung Bit 0 3 7 Nicht verwendet Octet 22 Supported warnings in Octet 20 Octet 23 Supported warnings...

Page 57: ...a gsd Hinweise zur gsd Datei in Abschnitt 12 1 beachten Siemens 6 Hardware Katalog aktualisieren Siemens 7 DME5000 PROFIBUS im Hardware Katalog PROFIBUS DP weitere Feldger te Encoder DME5000 PROFIBUS...

Page 58: ...inen Laptop mit serieller RS 232 Schnittstelle Im Lieferumfang des Schnittstellenadapters sind enthalten Schnittstellenadapter Steckernetzteil 24 V Verbindungskabel Laptop Schnittstellenadapter 9 poli...

Page 59: ...leepmodus Mit der Funktion Sleepmodus kann der Messlaser des DME5000 ber den Eingang MF1 aus und eingeschaltet werden Bei ausgeschaltetem Laser ist das DME im Standby Betrieb die Ready Anzeige im Disp...

Page 60: ...egel dB in Hex Format STX 0123 ETX Innentemperatur C in Hex Format STX 0126 ETX Geschwindigkeit x 0 1 m s in Hex Format STX 0135 ETX Service Status gem Display Icon 1 Icon aktiv 0 Icon inaktiv STX 012...

Page 61: ...1 STX 8135 4 0 F ETX Beispiel 2 8 m s STX 8135001C ETX Anforderung Service Status zum DME STX 0125 ETX vom DME STX 8125 2 0 F ETX Beispiel MF2 u PLB sichtbar in Display STX 812582 ETX 0x82 10000010 MF...

Page 62: ...00 als DeviceNet Slave wird anhand der Allen Bradley Software RS Networx for DeviceNet Revision 4 12 gezeigt Vorgehensweise DME5000 xx4 8 poligen M16 Ger testecker aufstecken und anschlie en 5 poligen...

Page 63: ...nts Siehe DeviceNet Specification 6 Offset Value 0 counts 3 6 7 Resolution 100 m 3 5 8 Serial Mode Baudrate 19200 3 2 1 9 Serial Mode Data 8n1 3 2 2 10 Plausibility Off 3 7 11 MF1 ActiveState High 3 3...

Page 64: ...DeviceNet Stecker ausger s tet Die Multifunktionsein und ausg nge MF1 und MF2 sind hardwarem ig nicht verf g bar Die Parametrierung von MF1 und MF2 ist trotzdem m glich und das Ergebnis kann im Diagno...

Page 65: ...Value High Byte Attribute 10 4 Laser on 1 Laser off 0 MF1 active 1 MF1 not active 0 MF2 active 1 MF2 not active 0 Reserved by DeviceNet Warning 1 No warning 0 Alarm 1 No Alarm 0 7 6 5 4 3 2 1 0 Warnu...

Page 66: ...Daten zyklisch gesendet oder wenn der Wert des Parameters Delta berschritten wird 13 7 6 Parameter Offset und Preset Der Parameter Offset wird bei Ausl sen der Presetfunktion berschrieben Funktionswe...

Page 67: ...el 13 DME5000 8009813 ZJV1 2017 05 SICK AG Deutschland Irrt mer und nderungen vorbehalten 67 Anhang 13 7 7 Sleepmodus Laser aus Die Funktion Sleepmodus kann mit dem Class Instance Editor aktiviert wer...

Page 68: ...chtig gespeichert Mit dem Class Instance Editor k nnen die Parameter nichtfl chtig permanent im DME5000 gespeichert werden Vorgehensweise siehe Screenshot Description Save Class 65 Instance 1 Execute...

Page 69: ...Fehlerstatus Description Get single Attribute Class 64 Instance 11 Attribute 12 Receive Data Beispiel 09 x 0 0000 1001 MF2 MF1 TMP LSR BUS ATT PLB RDY 0 0 0 0 1 0 0 1 Icob Bus und RDY sind im Display...

Page 70: ...escheibe die in starrer Verbindung mit der Motorwelle steht erzeugt Aus regelungstechnischer Sicht ergibt sich damit ein hoch dynamisches reines P Verhalten mit sehr rauscharmen sin cos Signalen Zus t...

Page 71: ...tion lesen 43h Position setzen 44h Analogwert lesen Temperatur C 46h Z hler lesen 47h Z hler erh hen 49h Z hler l schen 4Ah Daten lesen 4Bh Daten speichern 4Ch Status eines Datenfeldes ermitteln 4Dh D...

Page 72: ...Fehlern der keine Distanzwertberechnung mehr zulassen z B Plausibilit ts fehler infolge Lichtstrahlunterbrechung werden die analogen sin cos Signale auf die Nulllinie gezogen Ist der Lichtweg wieder...

Page 73: ...BUS Termination 85 6 Operation 86 6 1 Operating Panel Setup 86 6 2 Operation 86 6 3 Operating Panel 86 7 Menu Structure 89 7 1 Flow Chart 89 7 2 Explanations on the Menu Structure 89 8 Example 99 8 1...

Page 74: ...29 13 6 RS 422 Interface 129 13 6 1 Protocol 130 13 6 2 Commands 130 13 6 3 Examples for Commands Standard Protocol 131 13 7 DeviceNet Interface 132 13 7 1 General 132 13 7 2 Configuration 132 13 7 3...

Page 75: ...section 9 1 Dimensional Drawing Notes explain benefits of certain settings and support you in making best use of the DME5000 Warning notes Read and observe with care Warning notes protect from danger...

Page 76: ...Germany Subject to change without notice 8009813 ZJV1 2017 05 Intended Use 2 Intended Use The DME5000 is an opto electronic sensor used to record distances to a reflective tape that is moved linearly...

Page 77: ...chinery directive Observe the national safety and accident prevention provisions Repairs must only be performed by the manufacturer Manipulations and modifications of the device are not permissible Pe...

Page 78: ...he laser beam The DME receives the light reflected by the reflector and determines the distance between sensor and reflector by time of flight measurement The measured distance is submitted to a contr...

Page 79: ...nment in X and Y directions Select the reflector size so that the light spot does not move off of the reflector in case of vibrations If the reflector is attached to the moving part a smaller reflecti...

Page 80: ...pot to remain at the reflector center The signal damping displayed in display information field 1 indicates the current reception signal On clean surfaces the damping values listed in the table should...

Page 81: ...t 5 1 3 Alignment of Distance Measuring Device with Adjacent Data Transmission For ISD300 ISD400 1xxx ISD400 6xxx optical data transmission a minimum beam distance of 100 mm independently of the maxim...

Page 82: ...e 8009813 ZJV1 2017 05 Commissioning 5 2 Electrical Connection Wire DME5000 according to connection diagram Observe wiring notes see section 5 2 1 For plugs and cables see section 9 2 2 Fig 6 Connecti...

Page 83: ...E5000 8009813 ZJV1 2017 05 SICK AG Germany Subject to change without notice 83 Commissioning Fig 8 Connection diagram DME5000 xx4 DeviceNet Fig 9 Connection diagram DME5000 xx5 DeviceNet Fig 10 Connec...

Page 84: ...ss section Potential balancing currents through the cable shield must be prevented by suitable measures Interference free data transmission is only given when using shielded cables with twisted pair w...

Page 85: ...ect to change without notice 85 Commissioning 5 2 2 PROFIBUS Termination Fig 14 No pigtails connect shield short and over the entire surface ground BOTH sides Fig 15 Shield connection in plastic casin...

Page 86: ...functions 6 2 Operation The DME5000 is equipped with a membrane keyboard Only operate the input area with your fingers or a suitable pointer Never use any pointed or hard objects They may damage the...

Page 87: ...re failure message Reception level within the threshold range Reception level OK LSR Laser Pre failure measuring laser Laser OK PLB Plausibility Measuring error Cause e g light beam interruption conta...

Page 88: ...3 3 1 Entered parameter is invalid previous valid parameter remains saved Cause e g lower switching limit upper switching limit Remedy Correct parameters Information field 2 example 010 000 m Current...

Page 89: ...the field of view of the sensor No bus communication must be active Bus address changes will only become active after restarting the device If code input is skipped or a different figure is entered th...

Page 90: ...OK 1 Plausibility error 0 OK 1 Plausibility error Activation On default SSI interface in operation RS 422 interface not in operation Off RS 422 interface in operation SSI interface not in operation Th...

Page 91: ...le with Siemens Step 7 see chapter 13 Annex Hiperface BusAddress Add 00 default Setting of the Hiperface address area 0 31 Baud rate 0 6 kBd Baud rate setting 1 2 kBd 2 4 kBd 4 8 kBd 9 6 kBd default 1...

Page 92: ...ording to set baud rate Protocol Standard STX 8122 sign 7 0 9 ETX CRLF sign 7 0 9 CR LF CP0 sign 7 0 9 CP1 Not assigned For detailed description see section 13 6 RS 422 Interface DeviceNet Baud rate 1...

Page 93: ...ce default MF1 is used as distance switching output see 3 3 3 Service MF1 is used as service output see 3 3 4 Sleep mode MF1 is used for activation of sleep mode laser off confirmation by ready status...

Page 94: ...l device temperature upper set temperature limit Internal device temperature upper set temperature limit Off LowerTemp On default Undertemperature monitoring Internal device temperature lower set temp...

Page 95: ...ng in movement direction decreasing distance MF2 Parameterizable output Active State Parameterization of the switching logic Active 1 HIGH level at active output closer NO Active 0 LOW level at active...

Page 96: ...perature upper set temperature limit Internal device temperature upper set temperature limit Off LowerTemp On default Undertemperature monitoring Internal device temperature lower set temperature limi...

Page 97: ...ce version Offset The offset value is added to the internally determined distance value and acts on all outputs and the display Adjustable 250 m 0 is output via the SSI interface at a negative measure...

Page 98: ...Chapter 7 Operating Instructions DME5000 98 SICK AG Germany Subject to change without notice 8009813 ZJV1 2017 05 Menu Structure Reset NO YES No reset Reset to default values...

Page 99: ...parameters can be set this way During parameter change the reflector must be within the field of view of the sensor Step 1 Apply operating voltage Display Step 2 Access the menu CODE with Display Ste...

Page 100: ...2017 05 Example Step 5 Move forward a digit with and enter the desired 2nd figure with e g 1 Display Step 6 Move forward a digit with and enter the desired 3rd figure with e g 4 Display Step 7 Push t...

Page 101: ...awing DME5000 xx1 DME5000 xx3 1 x M16 plug SSI RS 422 DME5000 xx2 1 x M16 2 x M12 plug PROFIBUS DME5000 xx4 DME5000 xx7 1 x M16 1 x M12 plug DeviceNet 2 x M12 plug Hiperface DME5000 xx5 1 x M12 plug D...

Page 102: ...1 2017 05 Technical Data 9 2 Accessories 9 2 1 Reflectors Reflector 0 3 x 0 3 m 2 Diamond Grade mounted 1 Reflector 0 6 x 0 6 m 2 Diamond Grade mounted 1 Type PL240DG Part no 1017910 Type PL560DG Part...

Page 103: ...64 1 Type PL240DG H Part no 1022926 Type PL560DG H Part no 1023888 Reflective tape Diamond Grade size customizable Reflective tape Diamond Grade curve 749 x 914 mm 2 Type REF DG Part no 4019634 Type R...

Page 104: ...6032903 Cable socket M16 8 pin angle 5 m shielded Cable socket M16 8 pin angle 10 m shielded Type DOL 1608 W05MA Part no 2026743 Type DOL 1608 W10MA Part no 2027194 Cable 2 x 0 5 mm 2 6 x 0 25 mm 2 t...

Page 105: ...d PROFIBUS cable 2 x 0 34 mm 2 yard goods Type PR STE 1205 G Part no 6021354 Type LTG 2102 MW Part no 6021355 Temperature range moved 5 80 C Fixed 40 80 C Sleeve PUR violet 8 mm Shield AL PT foil Bus...

Page 106: ...GA Part no 6027533 Cable socket M12 5 pin straight Cable 6 m drop cable A coded Connection cable 6 m socket M12 5 pin straight plug M12 5 pin straight drop cable A coded shield to pin 1 Type DOL 1205...

Page 107: ...able plug 7 8 5 pin straight PG16 customizable Type DOS 7805 GK Part no 6028331 Type STE 7805 GK Part no 6028332 cable 2 x 0 34 mm 2 V V or 2 x 0 25 mm 2 CANH CANL twisted pair wires and shielded with...

Page 108: ...ded Cable socket M12 5 pin PUR straight shield on knurled nut A coded 2 m DOL 1208 G02MAH1 Part no 6032448 5 m DOL 1205 G05MAC Part no 6036384 5 m DOL 1208 G05MAH1 Part no 6032449 10 m DOL 1205 G10MAC...

Page 109: ...e without notice 109 Technical Data 9 2 3 Attachment Method Alignment bracket DME5000 stainless steel 1 4571 Type BEF AH DME5 Part no 2027721 9 2 4 Weather Protection hood Weather protection hood DME5...

Page 110: ...E5000 110 SICK AG Germany Subject to change without notice 8009813 ZJV1 2017 05 Technical Data 9 2 5 Passive Reflector 90 deflection mirror Type USP DME5 Part no 2027710 9 2 6 Interface Adapter Interf...

Page 111: ...ling casing for DME5000 Technical Data Part no 6036180 Casing material Glass fiber reinforced plastic GFK Operating ambience temperature 20 75 C short term 80 C Supply voltage DC 24 V 20 Power intake...

Page 112: ...D R H Service life at 25 C MTTF 50 000 h Light spot diameter distance max 130 mm 70 m Switching outputs MF1 MF2 B push pull Output MF1 MF2 HIGH UV 3 V LOW 2 V Input MF1 4 HIGH 12 V LOW 3 V Output curr...

Page 113: ...life at 25 C MTTF 50 000 h Light spot diameter distance max 270 mm 150 m Switching outputs MF1 MF2 B push pull Output MF1 MF2 HIGH UV 3 V LOW 2 V Input MF1 4 HIGH 12 V LOW 3 V Output current 5 100 mA...

Page 114: ...life at 25 C MTTF 50 000 h Light spot diameter distance max 550 mm 300 m Switching outputs MF1 MF2 B push pull Output MF1 MF2 HIGH UV 3 V LOW 2 V Input MF1 4 HIGH 12 V LOW 3 V Output current 5 100 mA...

Page 115: ...5000 8009813 ZJV1 2017 05 SICK AG Germany Subject to change without notice 115 Maintenance 10 Maintenance The DME5000 is maintenance free We recommend to clean the lens and reflector check screw and p...

Page 116: ...replacement device ready ATT display 32H Pre failure message contamination The measured values are still OK clean optical surfaces reflector lens TMP display 1E Internal device temperature has reached...

Page 117: ...g change bus communication was active there was no visual contact with the reflector Remove bus cable Check light path PROFIBUS error messages Problem Cause Countermeasure Device error Hardware proble...

Page 118: ...ermeasure Pre failure recognition Laser diode at the end of its service life Keep replacement ready for next maintenance cycle Light path is damped by mist or dust Check light path Lens or reflector c...

Page 119: ...peaters must be used to connect the bus segments The max cable length depends on the transmission speed see table 2 The indicated cable length can be increased by using repeaters We recommend not usin...

Page 120: ...ed onto the DME transmission line starting with the highest bit Between two cycle sequences there is a cycle break of at least 30 s The bit cycle is between 70 kHz and 500 kHz and depends on the line...

Page 121: ...00 kBd Trunk cable 500 m 250 m 100 m Drop cable 6 m 6 m 6 m Crop cable total length 156 m 78 m 39 m 12 5 Hiperface Hiperface means High Performance Interface and is a standard interface for motor feed...

Page 122: ...a a RS 422 command 2 Static preset Triggered via multifunctional input MF1 version DME5000 xx2 when using the GSD module Class 2 4 byte in out Procedure Set GSD parameter Preset mode to Preset mode Se...

Page 123: ...ending on device version DME5000 xx1 xx4 xx7 MF1 DME5000 xx3 MF1 or RS 422 command DME5000 xx2 MF1 or PROFIBUS command Function Triggering the preset causes internal calculation of an offset value tha...

Page 124: ...and offset with the preset mode parameters set SICK Class 1 See ENCODER Class 1 Class 2 4 byte input 4 byte output data device specific parameters are accessible via GSD and overwritten parameters ent...

Page 125: ...cording to resolution Value range 250 m 13 2 3 Diagnosis Data for Activated Parameter Extended Diagnostics DDLM_Slave_Diag Encoder profile Diagnosis Class 1 Octet 1 16 Octet 7 16 Class 2 Octet 1 63 Oc...

Page 126: ...Bit 0 Laser pre failure message Bit 1 7 Not used Octet 21 Warnings Bit 1 Device inside temperature Bit 2 Contamination Bit 0 3 7 Not used Octet 22 Supported warnings in Octet 20 Octet 23 Supported war...

Page 127: ...py GSD file in the gsd directory siemes step7 s7data gsd observe notes on gsd file in section 12 1 Siemens 6 Update hardware catalogue Siemens 7 DME5000 PROFIBUS in the hardware catalogue PROFIBUS DP...

Page 128: ...top with serial RS 232 interface The following is included in the interface adapter scope of delivery Interface adapter Plug in mains adapter 24 V Connection cable laptop interface adapter 9 pin Sub D...

Page 129: ...ptop via the monitor interface MF1 MF2 13 5 Sleep Mode The sleep mode function can be used to switch the DME5000 laser off and on via input MF1 With the laser off the DME is in standby operation the r...

Page 130: ...X 0122 ETX Reception level dB in hex format STX 0123 ETX Internal temperature C in hex format STX 0126 ETX Speed x 0 1 m s in hex format STX 0135 ETX Service status according to display icon 1 Icon ac...

Page 131: ...the DME 1 STX 8135 4 0 F ETX example 2 8 m s STX 8135001C ETX Request service status for the DME STX 0125 ETX from the DME STX 8125 2 0 F ETX example MF2 a PLB visible in display STX 812582 ETX 0x82...

Page 132: ...ation Configuration and commissioning of the DME5000 as a DeviceNet slave is explained based on an Allen Bradley software RS Networx for DeviceNet Revision 4 12 Procedure DME5000 xx4 Plug on and conne...

Page 133: ...viceNet Specification 6 Offset Value 0 counts 3 6 7 Resolution 100 m 3 5 8 Serial mode baud rate 19200 3 2 1 9 Serial Mode Data 8n1 3 2 2 10 Plausibility Off 3 7 11 MF1 ActiveState High 3 3 1 12 MF1 F...

Page 134: ...equipped with a 5 pin DeviceNet plug The multifunction in and outputs MF1 and MF2 are not available in the hardware Parameters for MF1 and MF2 can be set anyway and the result can be requested in the...

Page 135: ...Attribute 10 4 Laser on 1 Laser off 0 MF1 active 1 MF1 not active 0 MF2 active 1 MF2 not active 0 Reserved by DeviceNet Warning 1 No warning 0 Alarm 1 No Alarm 0 7 6 5 4 3 2 1 0 Warning All pre failu...

Page 136: ...Delta data are sent cyclically or if the value of the parameter delta is exceeded 13 7 6 Parameter Offset and Preset The parameter offset is overwritten when the preset function is triggered For prese...

Page 137: ...ns Chapter 13 DME5000 8009813 ZJV1 2017 05 SICK AG Germany Subject to change without notice 137 Annex 13 7 7 Sleep Mode Laser Off The sleep mode function can be activated with the Class Instance Edito...

Page 138: ...s them there volatile The Class Instance Editor can be used to permanently store the parameters in the DME5000 for procedure see screenshot Description Save Class 65 Instance 1 Execute message Executi...

Page 139: ...Error status Description Get single Attribute Class 64 Instance 11 Attribute 12 Receive Data Example 09 x 0 0000 1001 MF2 MF1 TMP LSR BUS ATT PLB RDY 0 0 0 0 1 0 0 1 Icob Bus and RDY are visible in th...

Page 140: ...cos signals are directly generated by scanning of the code disc inside the encoder which is firmly connected to the motor shaft From a control technical view this leads to a highly dynamic pure P beh...

Page 141: ...ion Code 0 1 Comment 42h Read position 43h Set position 44h Read analog value Temperature C 46h Read counter 47h Increase counter 49h Delete counter 4Ah Read data 4Bh Save data 4Ch Determine status of...

Page 142: ...ors that no longer permit distance value calculation e g plausibility error due to light beam interruption the analog sin cos signals are drawn to the zero line Once the light path is free again the z...

Page 143: ...22 deaktiviert SSI activated RS 422 deactivated 3 Nicht w hlbar wird nur angezeigt wenn SSI Monitor Mode ausgew hlt ist Not selectable only displayed if SSI Monitor Mode is selected 4 125 m bei Varian...

Page 144: ...t LowerLimit Service Laser Level UpperTemp LowerTemp Plausib ility Ready Bus Status Enter Value 002000 mm Enter Value 001990 mm Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set 3 3 1 3 3 2...

Page 145: ...utomatische Schnittstellenwahl je nach Ha Hinweis Parameter nderungen die mit best tigt werden sind sofort aktiv und nichtfl chtig gespeichert Hierbei muss der Reflektor im Sichtfeld des Sensors sein...

Page 146: ...il support sick jp Malaysia Phone 6 03 8080 7425 E Mail enquiry my sick com Mexico Phone 52 472 748 9451 E Mail mario garcia sick com Netherlands Phone 31 30 2044 000 E Mail info sick nl New Zealand P...

Reviews: