Chapter
13
Operating Instructions
DME5000
130
© SICK AG • Germany • Subject to change without notice
8009813/ZJV1/2017-05
Annex
13.6.1
Protocol
Depending on the protocol selected, the DME5000 transfers measured values as
displayed below:
Protocol
Standard
<STX>8122<[sign]><7*[0...9]><ETX> (request) or
<STX>0322<[sign]><7*[0...9]><ETX> (continuous)
CRLF
<[sign]><7*[0...9]><CR><LF>
CPO
<[sign]><7*[0...9]>
13.6.2
Commands
Request measured value/status/speed
Measured value (resolution 0.1 mm fixed)
<STX>0122<ETX>
Reception level (dB) in hex format
<STX>0123<ETX>
Internal temperature (°C) in hex format
<STX>0126<ETX>
Speed (x 0.1 m/s) in hex format
<STX>0135<ETX>
Service status according to display icon
1: Icon active, 0: Icon inactive
<STX>0125<ETX>
MF2 MF1 TMP LSR BUS ATT
PLB RDY
Continuous measured value output
Continuous output on
(resolution 0.1 mm fixed)
<STX>052201<ETX>
Continuous output off
<STX>052200<ETX>
Continuous speed output
Continuous output of a
speed (x 0.1 m/s) in hex format
<STX>053501<ETX>
Continuous output off
<STX>053500<ETX>
Special functions
Laser off (sleep mode on)
<STX>0333<ETX>
Laser on (Sleep mode off)
<STX>0332<ETX>
Preset activation
<STX>0335<ETX>