- 14 -
Table 5
Table 5
Table 5
Table 5-
-
-
-1 Settable
1 Settable
1 Settable
1 Settable parameters
parameters
parameters
parameters
No.
Parameter
Selectable value
Unit
Default
1
REC. RANGE
0.125 , 0.25 , 0.5 , 1 , 2 , 4 , 8 , 16 ,
32 , 64 , 128 , 256 ( 12 Steps )
μRIU/10mV
256
2
INTEG. RANGE
64 , 256
μRIU/1024mV
256
3
TEMPERATURE
OFF,30~55 (1 Step)
℃
40
4
TIME CONSTANT
0.1,0.25,0.5,1,1.5,2,3,6
(8 Steps)
sec
1.5
5
POLARITY
+,-
-
+
6
BASELINE SHIFT
0~200 (1 Step 50nRIU)
-
0
7
S.SEQ MODE
FINE, STANDARD, COARSE,
CUSTOM
-
STANDARD
8
LEAK SENSOR
ON,OFF
-
ON
9
KEY LOCK
YES,NO
-
NO
10 IP ADDRESS
0.0.0.0 ~ 255.255.255.255
-
192.168.1.2
11 SUBNETMASK
128.0.0.0 ~ 255.255.255.254
-
255.255.255.0
12
USE DEFAULT
GATEWAY
YES,NO
-
NO
13 DEFAULT GATEWAY
0.0.0.0 ~ 255.255.255.255
192.168.1.1
14 DEFAULT DATA
YES,NO
-
NO