Judge by Deviation Values
8.5.1
Set two limit values by putting actual samples on the balance
1 Start the limit function.
Press the Set key for a few seconds.
When [
] is displayed, release the key.
The current reference weight blinks in the display.
2 Save the reference weight.
Put a sample that provides a reference weight on the
balance, and press the Function key.
When a reference weight is saved, that value is
displayed for a while.
3 Save a lower limit.
The display first indicates [
], and then the
current lower limit blinks.
Put a sample that provides a lower limit on the
balance, and press the Function key.
A lower limit is saved. Then go to the next step.
4 Save an upper limit.
The display first indicates [
], and then the
current upper limit blinks.
Put a sample that provides an upper limit on the
balance, and press the Function key.
The upper limit is saved, and the balance goes back
into measurement mode.
* For the case of three or four point setting, repeat the above steps 3 and 4.
Limit values saved are displayed as [
] - [
] or [
] rather than
[
] or [
]. In addition, a triangle [
] on the left side of the display will be lit at
the same time for indicating the setting level.