21
5 Put additional samples on the balance.
Add no more than two times as many samples as the
displayed number and wait until it is stable again. A
beep will sound and the memory is updated.
Repeating this operation will improve the resolution of
parts counting, resulting in more accurate
measurements.
5 Save the average unit weight of the samples.
Press the Function key to complete the update of the
memory.
The balance beeps to indicate that the average unit
weight is saved, and then goes back into the
measurement mode.
Hints
1.
The [
] display that may be shown during a memory update of average unit weight
indicates that you added more than two times as many samples as the displayed number and
thus the counting accuracy is poor. Do not add too many samples at a time. Gradually
increase the number of samples to increase counting accuracy.
2.
The [
] display during a memory update of average unit weight indicates that the number
of samples you added is too small and the counting accuracy is poor. Add samples until the
[
] display is turned off to improve the counting accuracy.
3.
Even when these indications are displayed, you can press the Function key at any time to
store the average unit weight. In this case, however, counting accuracy is low, and a counting
error may be caused.
4.
When [
] is displayed, it indicates that the average unit weight is too small for the
balance to count the samples. For more information on the minimum unit weight supported by
the parts counting function, refer to Section 17: Specifications on page 73.
Pcs
Pcs
Function