8 Controlling and adjustment functions
FS series operation manual
-
70
-
(1) List of weights used for the span adjustment by model
Model name
FS623
FS3202
FS6202
FS15001
FS30K0.1G
Selectable
weight
620 g
3200 g
6200 g
15000 g
30000 g
600 g
3000 g
6000 g
500 g
2000 g
5000 g
10000 g
20000 g
200 g
1000 g
2000 g
5000 g
10000 g
100 g
500 g
1000 g
2000 g
5000 g
10 g
50 g
100 g
2000 g
500 g
“USER IN”
selection
0.001 g
-
620.000 g
0.01 g
-
3200.00 g
0.01 g
-
6200.00 g
0.1 g
-
15000.0 g
0.1 g
-
30000.0 g
Model name
FS60K0.1G
FS100K1G
FS200K1G
FS150K1GF
FS300K1GF
Selectable
weight
60000 g
100000 g
200000 g
150000 g
300000 g
50000 g
100000 g
200000 g
20000 g
50000 g
100000 g
50000 g
100000 g
10000 g
20000 g
20000 g
20000 g
50000 g
1000 g
2000 g
2000 g
2000 g
5000 g
“USER IN”
selection
0.1 g
-
60000.0 g
1 g
-
100000 g
1 g
-
200000 g
1 g
- 150000 g
1 g
- 300000 g
(2) “PUSH F” indicates models with a weighing capacity of not less than 60 kg.
(3) The span adjustment by the use of a weight less than the weighing capacity may possibly
indicate “UC” on the main LCD. When this is the case, the weighing accuracy is not
guaranteed.
Conditions under which “UC” is indicated
When an object that is more than two times heavier than the weight that was used for the
span adjustment is weighed, and
When the minimum indication setting (“61D. DA.”), which is finer than the minimum
indication setting (“SEL. DA.”) selected for the span adjustment, is performed.
8-17
Setting for maintenance
Setting menu “622.C2. to 625.R5.” are for the purpose of service maintenance. Please refrain
from performing setting.
If you should have performed setting, please notify the store where you purchased the
product.
Summary of Contents for FS100K1G
Page 2: ...白紙 ...
Page 95: ...FS series operation manual Appendix 85 Appendix 1 3 Dimensional outline drawing ...
Page 101: ...FS series operation manual Appendix 91 Hierarchy of functions related to the lock ...
Page 106: ......
Page 107: ......
Page 108: ......