6 External input / output functions
FS series operation manual
-
48
-
6-4-4
Input command format
Composed of four characters including a terminator (CR=0DH / LF=0AH).
1
2
3
4
C1
C2
CR
LF
6-4-5
Transmission procedure
1
Send an input command from an external device to the scale.
Since transmission and reception are performed by way of full-duplex communication method, the
input command can be transmitted irrespective of the transmission timing from the scale.
2
When the scale has successfully executed the input command received, the scale sends a normal
response or the data requested by the input command. In the case of unsuccessful completion or
reception of an invalid input command (error), the scale sends an error response. In the normal
operation, the scale normally sends a response within one second after an input command is
transmitted.
However, the response is sent after completion of the processing when:
(1) A tare weight subtraction command or a zero adjustment command is received when the
setting menu is set to the “17 .TA. 1 Stability waiting”, or
(2) It takes time to process the input command received.
In addition, input commands received in other than the scale mode are neglected.
After transmitting an input command from an external device, please do not send a next
input command till receiving a response from the scale.