background image

7

Op e r a t i on (con t i n ue d)

Ch a n g i n g Memo r y M odes

The DRI operates in three distinct memory modes:

• Single (on-demand memory)
• Continuous
• Standard  

(

NOTE

: Main display reflects “nonE” in this mode)

To change the memory mode:

1. Press the PEAK button while holding down the ON button; the small display will reflect “dSP”.
2. Press the PEAK button six times; “Log” will appear on the small display and the main display will reflect the current setting.
3. Each time the > button is pressed, the main display will scroll through “SinG”, “Cont” and “nonE”.
4. Press the MEMO SET button to store the desired memory mode and exit.

NOTE

: Data stored in memory can be output to an external device (see “Communication with External Devices” section).

Single and Continuous Memory Modes

The three functions performed when using either the Single or Continuous mode are:

• Storing data
• Recalling data
• Deleting data

Storing Data (Single Memory Mode)

After selecting the “SinG” mode it is possible to send one data reading into memory each time the MEMO SET button is pressed; the
small display will briefly change to an “M”.  The “M” indicates that the data reading displayed has been stored in memory.  Up to one
hundred data readings can be stored. “FULL” will appear on the main display if memory is full.

Storing Data (Continuous Memory Mode)

After selecting the “Cont” mode it is possible to enter data readings continuously into memory by pressing the MEMO SET button once.
The small display will reflect a flashing “M” and data readings will be stored in memory as quickly as the display update rate chosen
(reference “Changing Display Update Time” section).  Data readings will continually be entered into memory until the MEMO SET
button is pressed a second time or 100 samples have already been entered (“FULL” will appear briefly on the main display).

Recalling Data (Single and Continuous Memory Mode)

To recall data in the memory press the RECALL button; the last data reading entered and corresponding memory position will be flashing
opposite each other on the main display.  Each subsequent push of the RECALL button will scroll through the data readings in the
opposite order as they were entered.  Press the RESET button to exit from memory recall.

NOTE

: No data readings have been entered if “nonE” shows on the main display after pressing the RECALL button.

Deleting Data (Single and Continuous Memory Mode)

While recalling data (see above) it is possible to delete one data reading at a time or delete the entire memory.
To cancel one reading at a time:
Press the RECALL button to scroll through the data in memory until the undesired reading is appears on the display.  Push the > button to
delete that data reading.  All higher memory positions will decrement one to fill the deleted position.  If the reading being deleted is the

last one in memory, then the memory position will decrement one as there is no higher memory position to take its place.
To delete the entire memory:
After recalling the memory using the RECALL button, press the RESET button while holding down the ON button; “CLr” will briefly
appear on the main display.  This indicates that all data readings have been deleted and will automatically exit from memory recall.

Summary of Contents for DRI

Page 1: ...al Remote Indicator with Remote Load Cell Torque Sensor Inst ruction Manual Inspection Standard Accessories 2 Features and Benefits 2 Important Safety Instructions 2 Set Up 3 Power 3 Connection to Rem...

Page 2: ...ard or foot switch Do not use any chemicals to clean the case use a wet soapy cloth Do not use or store unit in oily dusty or excessively wet areas Do not use or store unit in extreme temperatures nor...

Page 3: ...plug with the arrows on the port of the DRI 3 Push the plug straight onto the port until a click is heard 4 Press the DRI s ON button the LCD will briefly display all indicators see Display Keypad se...

Page 4: ...y 5 Small Display 6 Maximum Reading Indicator 7 Minimum Reading Indicator 8 AC Adapter Port 9 Button 10 HI LO Button 11 Unit Button 12 Peak Button NOTE Refer to Changing the Measuring Mode section for...

Page 5: ...button will allow the display to scroll from average to peak compression or clockwise torque to peak tension or counterclockwise torque When in the Standard mode pressing the PEAK button will allow th...

Page 6: ...the small display and whatever limit is already programmed will appear on the main display 7 To program the low limit repeat steps 2 5 8 Once the low limit has been programmed press the MEMO SET butto...

Page 7: ...stored in memory as quickly as the display update rate chosen reference Changing Display Update Time section Data readings will continually be entered into memory until the MEMO SET button is pressed...

Page 8: ...of the Peak button will scroll to minimum peak peak average standard deviation and return to maximum 4 Press the RESET button to exit NOTE Peak and peak will not be displayed if the data was stored in...

Page 9: ...ng down the ON button The small display will indicate dSP 2 Press the PEAK button four times The small display will indicate OUT and the main display will reflect the current setting 3 Each time the b...

Page 10: ...the HI LO button the small display will indicate dEL and the main display will reflect the current setting 11 Each time the button is pressed the main display will toggle between Cr and CrLF 12 To set...

Page 11: ...riptions ASCII capital characters UNTG UNTN UNTL Change UNITS to Kg N lb UNTG CR Kg g UNTN CR N UNTL CR lb oz DSPN DSPT DSPC Change to average peak tension peak compression mode DSPN CR Change to aver...

Page 12: ...AK DATA DATC CR Output compression PEAK DATA AB CR Stop continuous data output BB CR Continuous data output BC CR Model name confirmation request BD CR Unit s confirmation request Typical gauge respon...

Page 13: ...seven times The small display will indicate dAf and the main display will reflect the current setting 3 Each time the button is pressed the main display will toggle between on and oFF 4 Press the MEMO...

Page 14: ...ressed the main display will scroll through tArE HoLd nonE 4 A If nonE or tArE is desired push the MEMO SET button to store the selection and exit B If the HoLd feature is desired push the HI LO butto...

Page 15: ...om an external source in the edge trigger mode To release it press RESET Level mode can be released externally Level trigger OPEN Display holds present value when switch opens Level trigger CLOSED Dis...

Page 16: ...rns ON there are two OC NPN transistors one for tension and one for compression The output of these transistors can be used as an alarm or to stop a process thus protecting the remote sensor or the sa...

Page 17: ...n of descriptions of the options within each parameter Parameter Settings DESCRIPTION DISPLA Y OPTIONS Display Update Time in seconds see page 6 0 125 0 25 0 5 1 2 Measuring Mode on oFF see page 5 Aut...

Page 18: ...sure the power on the DRI is turned OFF 4 Plug in the remote load cell to the DRI 5 While holding down the MEMO SET button press the ON button and hold until the main display shows CAL The small displ...

Page 19: ...lt to factory settings Memory data error Press RESET to clear memory Calibration error Press RESET and repeat calibration procedure Memory unit error occurs if the Change measurement units or clear me...

Page 20: ...the entire and final expression of the parties agreement War ranty Dimensions and Specifications S P E C I F I C A T I O N S MODEL MODEL MODEL MODEL MODEL DRI DRI DRI DRI DRI Sensor Sensor Sensor Sens...

Reviews: