CONNECTION OF NON-SHIMADZU BUCKY SYSTEM
L 100
24 V DC
24 V DC
K 27 B
K 31
* 1B22
BUCKY 1
* XOK
L 0
1 B11
1 B1
1 B22
1 B2
X 2
TERMINAL-LF30
TERMINAL-LFX
NON-SHIMADZU BUCKY
JD 2
X0 1D
X0 1E
9
8
1
2
K 03
L 0
L 100
9
10
1 B22X
1 B2X
1 B1X
1 B11X
4
1
3
2
JP 2C
JB 2B
JB 2A
CONFIRMATION OF
GRID MOVING
RETURN SIGNAL
POWER SUPPLY TO
DRIVE GRID MOTOR
GRID MOTOR
CAUSION: MAKE SURE THERE
HAS NO JUMPER CONNECTED HERE
•
Change jumper positions according to the voltage on 1 B11X.
JP 2C: AC 100-125 V
JP 2B: DC 24 V
JP 2A: DC 12 V
Jumper position of JP 2A, JP 2B, and JP 2C depends on the voltage of from 1 B11X, grid confirmation
signal provided by the Bucky system. If Bucky provide a 24 V DC signal to 1 B11X, when grid is
moving, install JP 2B to 1-4 position, and JP 2C, JP 2A to 2-3 position. Then connect 1 B1X to 24 V
DC ground. Make sure only one jumper is installed on 1-4 position and the other two on 2-3 position.
•
Connection of I. I. Mag. mode selection signal
I.I. Mag. mode signal is connected via a new cable, I.I. View-field switching cable (P/N 532-15191)
provided for this generator. This cable connects between JSK terminal on TERMINAL-LF30 PCB and
IS connector attached to I.I. (Refer to Circuit diagram SYSTEM-LF30 15/18)
UD150L-30/FX
YSF series table
TERMINAL-LF30 PCB
GENERATOR I/F PCB
JSK
IS
J9C
View-field switching cable
Adapter
UD150L30-FX/EX Installation Guide, REV_B Page 9 of 35