Fluoro IBS signal connection, case 2
When using TV signal to regulate fluoro IBS, a shielded co-axial cable should be connected from TV
camera to ANALOG-LF30 PCB.
UD150L-30/FX TV
system
ANALOG-LF30 PCB
TV 7
IBS signal
Shielded cable
TV 2
0 V
•
Photo timing connection
•
For SHIMADZU Bucky with SPT photo timing connection, check the installation manual section 6.2.
•
If you programmed the Comet Chambers to Detector 2, follow the stated below diagram for
connection.
Comet
Chamber
PH
CONT-98
PCB
1
2
3
4
1
2
3
4
SIGNAL
SIGNAL
CN2-DTA
PLUG
1
2
SIGNAL
CN2-DTB
in case of DET 2
1
2
Left
Field
Right
Field
Center
Field
Plug
2
5
3
4
1
6
You can change the connection anyway you like. Just make sure the connection stay in the same
group. For example,
CN1-DTA and CN1-DTB are for Detector 1
CN3-DTA and CN3-DTB are for Detector 3
UD150L30-FX/EX Installation Guide, REV_B Page 11 of 35