Bucky connection
CAUTION:
If UD150L30-FX is equipped only with SHIMADZU Bucky,
YOU MUST
DISCONNECT CABLES
between TERMINAL-LF30 and TERMINAL LFX PCB. This includes
1B11, 1B1, 1B22, 1B2, 2B11, 2B1, 2B22, and 2B2 terminals on TERMINAL-LF30 PCB. Tape all
disconnected terminals separately and secure them with other cables. Then connect Bucky cable
directly to TERMINAL-LF30 PCB as stated below. Fail to disconnect the cable between these two
boards will damage TERMINAL-LF30 PCB.
CONNECTION OF SHIMADZU BUCKY SYSTEM
TERMINAL-LF30 PCB
SHIMADZU BUCKY
L 100
X 2
24 V DC
L 100
L 100
* 1B 22
* XOK
BUCKY 1
L 0
K 25 A
K 27 B
K 31
24 V DC
1B11
1B1
1B22
1B2
Note 1
Note 2
B1
B2
L 100
L 0
GRID MOVING
CONFIRMATION
AC POWER TO
DRIVE BUCKY
GRID MOTOR
##
## Add Jumper
Note 1:
*1B22 is a DC signal within TERMINAL-LF30 PCB. When selecting Bucky1, 1B22 is at DC
ground. However, 1B22 at terminal X2 is a AC 100 volt power supply to drive grid motor in Bucky.
DO NOT mix these two signals by adding any extra jumpers for trouble shooting purpose.
Note 2:
*XOK is the signal to turn on X-ray. Can be found in SYSTEM-LF30 PCB print 10/18.
(M74A pin 2)
UD150L30-FX/EX Installation Guide, REV_B Page 8 of 35