X-ray tube data for Nonstandard Tube Data Mode
X-ray Tube Type /Focus size
Type
Large focus
Small focus
RAD-60 1.2
mm
0.6mm
Anode rotation
High speed rotation
Maximum KV
150 KV
Minimum KV
40 KV
HU load factor
90 %
CINE load factor
90 %
HU cooling curve
HU(KHU)
Time (min)
1
400
0
2
260
1
3
200
3
4
150
10
5
0
40
Emission characteristic
Large focus
Small focus
KV
mA
KV
mA
1
40
470
40
220
2
50
580
50
340
3
60
800
60
400
4
80
940
80
450
5
0
0
0
0
Short time raiting
Large focus
Small focus
KV
mA
KV
mA
Sec
1
110
580 100 225 0.01
2
110
570 100 225 0.03
3
110
540
100
220
0.1
4
100
500
100
200
0.3
5
100
360
80
220
1
6
80
300
80
160
3
7
80
270
80
100
10
8
0
0
0
0
20
9
0
0
0
0
50
If max
Large focus
Small focus
5.5
A
5.2
A
mA position(Recommended)
Large focus(mA)
Small focus(mA)
1
630
250
2
500
200
3
400
160
4
320
100
5
250
80
6
200
50
Large focus(mA)
Small focus(mA)
Maximum mA
800
320
Minimum mA
10
10
UD150L30-FX/EX Installation Guide, REV_B Page 31 of 35