X-ray tube data for Nonstandard Tube Data Mode
X-ray Tube Type /Focus size Type
Large focus
Small focus
RAD-56
1.2mm
0.6 mm
Anode rotation
High speed rotation
Maximum KV
150 KV
Minimum KV
40 KV
HU load factor
90 %
CINE load factor
90 %
HU cooling curve
HU(KHU)
Time (min)
1
400
0
2
260
1
3
200
3
4
150
10
5
0
40
Emission characteristic
Large focus
Small focus
KV
mA
KV
mA
1
40
420
40
120
2
50
600
50
160
3
60
800
60
200
4
80
940
80
230
5
0
0
0
0
Short time raiting
Large focus
Small focus
KV
mA
KV
mA
Sec
1
110
500
100
195
0.01
2
110
500
100
190
0.03
3
110
470
100
180
0.1
4
110
400
100
170
0.3
5
80
400
100
150
1
6
80
250
80
130
3
7
80
130
80
80
10
8
0
0
0
0
20
9
0
0
0
0
50
If max
Large focus
Small focus
5.5
A 4.8
A
mA position(Recommended)
Large focus(mA)
Small
focus(mA)
1
630
250
2
500
200
3
400
160
UD150L30-FX/EX Installation Guide, REV_B Page 32 of 35