TOCV-CP_IMshort_Control-ver-2_Rev07F
Page 14 of 22
(Software version ≥ 2.00)
effective date: September 2007
Expected Conc. Range
Enter the maximum expected concentration. A range value is
automatically entered based on the selected calibration curve
No. of Injections
Enter the minimum/maximum number of injections for the
samples
SD Max
CV Max
Enter the maximum standard deviation and coefficient of variance
that is acceptable for the injections. If one of these value is met, no
additional injections are required. If both values are exceeded, the
samples are automatically re-injected up to the maximum number
of injections.
No. of Washes
Enter the number of times the syringes is to be washed with
sample before the first analysis injection
Auto dilution
Enter the dilution factor for the samples. The instrument will
automatically dilute the sample by this factor, and the result will
be multiplied by the factor to obtain the final concentration.
Sparge Time
Enter the required sparge time. This option is available only for
NPOC analysis.
Acid Addition
Enter the percentage of acid to be added to the sample
Multiple Injections
If this item is selected, multiple injections are made from a single
sample aliquot. This larger aliquot is drawn in the syringe in one
aspiration and is sufficient to accommodate all of the repeat
injections
Use blank check area
If selected, the blank check value is subtracted from the analysis
result
Auto. Correction of inj.
Vol. and dilution
If selected the instrument will automatically adjust the injection
volume and dilution factor and conduct re-analysis when the
measurement result exceeds the calibration curve range
Method Wizard –5 Peak Time Parameters
Parameter
Action/Description
Analysis
Displays the selected analysis type
Use default settings
If this item is selected, the software default values are used for
peak detection and the Min. integration time and Max. integration
items are disabled
Min. integration time
Enter the minimum time the instrument signal will continue to be
detected
Max. integration
Enter the maximum time the instrument signal will be detected
when no peak is found
Note: Wizards from 3 to 5 are displayed for each single analysis type.
Click Finish and the method file is saved