TOCV-CP_IMshort_Control-ver-2_Rev07F
Page 19 of 22
(Software version ≥ 2.00)
effective date: September 2007
5.
Vial number
After inserting the samples in the sample table, the vial number of vials to be sampled I the
ASI-V must be associated with the samples
-
Open a complete sample table and click the Vial setting button
Parameter
Action/Description
Table
The sample table displays the row, sample name and attribute of the
inserted samples. Content in the Row, Sample Name and Attribute
columns cannot be edited
Vial arrangement
Designated vials are marked blue. If the total injection volume
exceed the capacity the vial is marked red
Vial
“!” is displayed if the total injection volume exceeds the vial
capacity
Acid addition
Select a vial in the drawing to display the percent of acid to be
added to the vial
Sparge Time
Select a vial in the drawing to display the length of time sparging
will be conducted in that vial
Required Sample Amt. Select a vial in the drawing to display the total injection volume
from that vial
Off-Line
Select a cell in the Vial and enter a “0” to conduct sampling from
the sampling tube on the left side of the instrument
[OK]
To verify the information. The Sample Table is displayed with the
inserted analysis
Enter the vial number by:
a) Keyboard entry
b) Entry by Dragging
c) Entry by double-clicking the Vial arrangement drawing