1 OUTLINE
MUX-100 SERVICE MANUAL 1 - 2
1-2 Specifications
・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・
MUX-100D
MUX-100DJ
Use environment
Atmospheric pressure
700
~
1060hPa
Ambient temperature
10
~
35
℃
Relative humidity
30
~
75%
Common specifications
Total width
580mm
Total length
1250mm
Height of support stand
1930mm
1780mm
Weight
39
4
kg (with imaging unit)
Maximum running speed
5
±
0.5km/h(this may vary depending on the floor condition.)
Imaging unit storage
1 Imaging unit and 1Gid
Focus height
600
~
2010mm 600
~
1860mm
Arm length
635
~
1200mm
Support stand rotation range
±
270deg min.
Tube rotation angle
360deg
Tube rotation around tube axis
Forward: 90 deg min.; Backward: 20 deg min.
Minimum distances between Focal
Spot and Image Receptor
312mm
Power supply
Supply voltage
Built-in battery
For battery drive:
Battery type
Shielded lead storage battery (12 V
×
20 cells)
Rating of power supply
1kVA
Supply voltage
Single phase AC 100,110,120,200,220,230,240V
Supply Impedance
Single phase AC 100,110,120 V : No more than 1.0
Ω
Single phase AC 200,220,230,240 V : No more than 4.0
Ω
For battery charge:
Supply frequency
50/60Hz
Power cable length
4m