![Shimadzu MUX-100D Service Manual Download Page 10](http://html1.mh-extra.com/html/shimadzu/mux-100d/mux-100d_service-manual_1256940010.webp)
1 OUTLINE
MUX-100 SERVICE MANUAL 1 - 7
CXDI-50G
CXDI-50G is containing mainly three componets;
Imaging unit; Digital Cassette that has the mobility and can be used on the optional angles
Power box; Power supply and data gateway for imaging unit mounted in the MUX-100D.
CXDI ; Digital Radiography Operation System for CXDI-50G, installed in Control PC and controled with Touch panel
monitor on the MUX-100D.
(1) Imaging
Unit
The sensor unit converts the X-ray image to the electrical signal (O/E Conversion) and after performs the A/D conversion,
transfer its signal through the Power Box with Ethernet cable to the Control PC.
Imaging unit
Effective range
353 x 430 mm
Number of Pixels
2214 x 2700
Effective Number of Pixels
2208 x2688
Pixel Pitch
160
µ
m x 160
µ
m
Fluorescent substance
GOS Fluorescent screen
Imaging Unit mass
4.8kg (except Grid)
Mechanical strength
Load uniformly: 150kg
Load partly: 100kg/
φ
40mm
Mass
4.8 kg (Except cable)
Attenuation equivalent of front panel
0.2mmAl.Eq.
Dimension
491(W) x 477(D) x 23(H)mm with operation grip
Power Box
Communication interface standard
IEEEE 802.3u (100BASE-TX)
Connector type: RJ45
Dimension
358(W) x 200(D) x 65(H)mm
CXDI
Image field sensing
Automatic field sensing
Grayscale
12bit (4,096 grayscale)
Preview image access time
Approx. 3 second
Data output
DICOM 3.0 compatible, Print Management Service Class (SCU), Storage
Service Class (SCU) (JPEG transfer syntax available)
Storage
Temporary storage in Hard disk of the Control PC
Option
Grid
Attenuation
equivalent:2.3mmAl.Eq.
For detail about CXDI, please refer to CXDI series operation manual (Canon CXDI Digital Radiography).