4 INSTALLATION AND MAINTENANCE/INSPECTION
MUX-100 SERVICE MANUAL
4 - 24
2) Any corrosion?*
inspection” on the
next page.
Bumper
Bumper switch
Do you operate normally?
Any loose parts?
Is the cover correctly attached?
Does the bumper switch operate normally?
Catcher part
Any unusual looseness to the top plate?
Earth belt
1) Does it appear (look) normal?
2) Is any debris attached to the belt?
Remove dust,
when dust has
adhered to the
belt.
Power plug
Is there any gap between the plug and the stopper?
Are any of the plug’s prongs bent or deformed?
Cord reel
Is anything caught in it and making noise?
Is the cable cover cut or frayed?
FPD box fan vent holes Is dust collected on the vents?
Is air being exhausted normally from the vents?
Others
Any presence of unusual smells or abnormal
sounds when the power is turned on or while the
battery is being charged?
Any abnormal sound or vibration coming from the
X-ray tube assembly?
Table 4-9 Daily check items
Please refer to Chapter 6.1 “Inspection” in Operation Manual of MUX-100D about the detail of daily inspection of each parts.