Trouble Shooting
Action
Message
Description
LX-Tube Water Level Warning
Water volume of heat
exchanger gets low
Contact Shimadzu
Service Representative.
(It
ʼ
s possible to do
exposure X-ray while this
Warning displays.)
LX Tube Thermal Warning
X-ray tube unit temperature
exceeded 70 celusius
degrees. The continued
use can be detect "Thermal
Over".
Cease operation or
decrease frequency of
exposure X-ray. (It
ʼ
s
possible to do exposure
X-ray while this Warning
displays.)
Starter is not working
The starter is in standby
status.
Contact Shimadzu service
representative.
Error Messages at Power ON
Action
Message
Description
Contact Shimadzu service
representative.
The data recording battery in
the control cabinet must be
replaced soon.
The data recording battery in
the control cabinet is empty.
Cabinet Battery Empty
Cabinet Battery Warning
Power OFF the high
voltage system, and reset
it after confirming the
switch of fluoroscopy and
radiography is not ON.
Radiography switch was
ON, when the high voltage
system is ON.
Fluoroscopy switch was
ON, when the high voltage
system is ON.
Fluo switch Error
X-ray switch Error