System Startup and Shutdown
Be careful!
During warm-up, X-ray exposure is performed automatically without operating
hand switch or foot switch.
CAUTION
The warm-up procedure can be performed when C-arm is at
the park position even though the "fluoroscopy at park position"
setting is set to OFF. In this setting X-ray exposure by the foot
switches and the hand switch is not available.
NOTE
X-ray Generator Console
1
Press
on the [Tool] menu
Warm-up window is displayed.
2
Press
Warm-up procedure is started.
The progress of warm-up is displayed.
When warm-up is complete, the progress bar
displays full.
X-ray Tube Warm-up