49
6
Trouble Shooting
1
Select and click on the
side menu
FPD calibration window appears.
2
Make sure that [Timeout
(Seconds)] is set to 10 seconds
This defines the maximum number of seconds
to wait before cancelling if there is a problem
communicating with the FPD.
3
Select [Offset-All Modes] in the
[Calibration Type] list.
4
Click
Progress messages are displayed during
calibration.
Completion message is displayed when
calibration is completed.
5
Click
Click
if you terminate the calibration before
completing.
If you terminate the calibration, be sure to calibrate again before
acquisition.
NOTE
FPD Problem
Automatic FPD calibration is supported to the system. Manual calibration is not necessary, so follow the
instruction below when manually calibrating the system.