OPERATION Of BL series 2/2
PCS SETTING
STEP
(1) Register the PCS unit following the unit registration.
(Refer to the following.)
When PCS unit has been already registered, it is not
necessary to register again.
(2) Place the tare on the pan and press the TARE key.
(3) Load the standard sample with required pieces.
(4) Check that the stability mark is lit.
(5) When pressing the UNIT key continuously, the
display will change as follows:
“
”, “
”,
“
”, and “
”
(6) Select the desired PCS display and press the TARE key.
(7) “ ” is displayed for several seconds and the balance enters the PCS
unit display. When the reference sample weight is less than “readability x
No. of set pieces”, the PCS setting is not made.
In this case, “ ” is displayed for several seconds and the balance
returns to weight display.
% SETTING
STEP
(1) Register the % unit. (Refer to the following.)
When % unit has been already registered, it is not
necessary to register again.
(2) Place the tare on the pan and press the TARE key.
(3) Load the reference sample.
(4) Continuously press the UNIT key to display
“
”.
(5) After the stability mark is lit, press the TARE key.
(6) “ ” is displayed for several seconds and the
balance enters the % unit display.
Minimum displayed value changes as follows depending on reference
sample weight (REF.)
Reference sample weight (REF.) is count value which minimum displayed
value rearranges one count.
If the % conversion is not possible, “ ” is displayed for several
seconds and the balance returns to weight display.
REGISTRATION, CANCEL, AND CHANGE OF UNIT
Step for registration
(1) Press the CAL/MENU key and select the
display.
(Press the TARE key.)
(2) Press the CAL/MENU key and select the
display.
(Press the TARE key.)
(3) The registrable unit is displayed by every pressing the CAL/MENU key.
The registerable units are three kinds which are selected from the following
14 kinds.
g, kg, ct. pcs, %, oz, ozt, dwt, GN, Hong-kong tail, Singapore tail, Taiwan
tail, Maraysia tail, and Japanese “monme”
However, % and pcs (No. of pieces) cannot be registered simultaneously.
The stability mark is lit on the unit display currently registered.
(4) Press the TARE key on the unit display to be registered. That unit is
registered.When three kinds of unit are already registered, a new registration
deletes the oldest registration among the three registered units. However, %
and pcs (No. of pieces) cannot be registered simultaneously. Then deletes
unnecessary one.
(5) Continuously pressing the TARE key returns the display to weight display.
Step for cancel
(1) Carry out the same operation described (1) to (3) above to set unit display.
Selecting the same one of the unit display which is currently registered
(stability mark is lit) cancels the registration.
Step for change
(1) Pressing the UNIT key changes the unit which is already registered.
However, even if the unit of % and pcs is already registered, the display does
not change to this unit unless making a setting of reference value.