Calibration
Calibration
Adjustment and Tests with Internal Weights (W-AD/ W/ X Series only)
[Internal Weight Adjustment]
Correctly adjusts the span of balance with the use of internal weights.
[Internal Weight Test]
Uses internal weights to verify if the sensitivity of the balance is accurate. *Does not calibrate
sensitivity.
First of all, register [Internal Weight Adjustment] or [Internal Weight Test] in [
CAL].
([Internal Weight Adjustment] is registered as the default.)
1.
Start adjustment or test
Press [
CAL].
The display will switch to an execution in program
screen when this is pressed, and adjustment or tests
with the use of internal weights will be started.
When Canceling Calibration...
Pressing the [
CAL] key or [
POWER] key during calibration will cancel the calibration.
Do not cancel calibration with the keys or weighing tasks when [Please wait a moment] is
displayed.
The screen will return to the weight measurement mode when adjustment and test are
complete.
Test Results
If any sensitivity deviations which exceed the inspection tolerances (daily inspection tolerances
specified by the customer) are found in the results of the calibration checks, carry out span
calibration.
CAUTION
Instructions
Do not leave or move the balance if adjustment or tests have not been
completed or ended normally
The internal weights are not maintained, so moving the balance may result in malfunctions. Make sure that the power is
switched off and on once so that the balance can be started up normally (with the internal weights maintained).
When performing adjustment with internal weights
When performing tests with internal weights
(reads the error of span)
*
*