Connections and Communications with External Equipment
Connections and Communications with External Equipment
4.
Set the zero return requirement in accordance
(1) Select [Zero Return Requirement], and then press [
OK].
(2) Select [Zero] or [50% of Previous Output], and then press [
OK].
What is the Zero Return Requirement?
Depending on the zero range setting, the word “zero” in zero return judgment means that if the
weighing value was less than the zero return requirement value when the previous sample was
removed, it will not be possible to place the next sample on the pan and automatically output
it until the display becomes stable. This function prevents more than two outputs for the same
sample. Select either [Zero] or [50% of Previous Output] for the zero return requirement value.
The 50% setting enables time to be saved by allowing the next sample output as soon as the
display stabilizes without completely returning to zero.
5.
Return to the weight measurement mode
Press [
POWER].
6.
Check the auto print function
(1) Place the container on the pan, and then press [
0/T].
(2) Place the sample in the container.
(3) Output will be confirmed automatically when
(Stability Mark) is displayed.
(4) Remove the sample from the pan.
*