163
12
12
Specifications
Specifications
1
3
5
2
4
6
7
8
12
14
13
9
10
11
Series Name
X Series
Y Series
Model
AP124X
AP224X
AP324X
AP124Y
AP224Y
AP324Y
Weighing Capacity
120g
220g
320g
120g
220g
320g
Minimum Display
0.1mg
Calibration Weight
Built-in
None
Range of External
Weights for
Calibration Purposes
(Recommended weight)
45~120.0090g
(100g)
95~220.0090g
(200g)
95~320.0090g
(300g)
45~120.0090g
(100g)
95~220.0090g
(200g)
95~320.0090g
(300g)
Repeatability
(when under weighing
capacity load)
0.1mg
0.15mg
0.1mg
0.15mg
Repeatability (when
under low load) (*1)
0.1mg
(load weight 5g)
0.1mg
(load weight 10g)
0.1mg
(load weight 20g)
0.1mg
(load weight 5g)
0.1mg
(load weight 10g)
0.1mg
(load weight 20g)
Minimum weighing
value (*1)
200mg
Linearity
±0.2mg
±0.3mg
±0.2mg
±0.3mg
Display response
time for small quantity
weighing (*2*3)
Approx. 2 sec.
Response Time (*3)
Approx. 2 sec.
Temperature/ Humidity
Range During Use
5~40°C, 20~85% (*4)
Sensitivity Temperature
Coefficient
±2ppm/°C (10~30°C)
Pan Size
Approximately 91DIA. mm
Main Unit Dimensions
Approximately 212(W) × 367(D) × 345(H)mm
Weight
Approximately 7.0 kg
Approximately 6.5 kg
Display
Organic EL Display (Dot Matrix)
Rated Power Supply
DC 12V , 1.0A
Degree of Soiling
2
Overvoltage Category
II
Altitude
Maximum 2,000m
Installation
Environment
Limited to Indoor Use
Input Power Supply
(AC Adapter)
AC100-240V, 0.32A 50/60Hz (*5)
I/O Terminals
RS232C (D-sub9P Plug), USB device (Type B), Ionizer
RS232C (D-sub9P Plug), USB device (Type B)
*1 USP41 compliant. Representative values from testing using weights equal to approximately 5% of the weighing capacity (or 5 g) at
the Shimadzu Corporation factory. The minimum display value will be affected by the environment in which the balance is installed,
so measurement needs to be carried out in the actual usage location.
*2 Time until 90% of the value of the sample quantity added in small quantity weighing (1 mg) is displayed.
*3 The response time is a representative value.
*4 No condensation.
*5 Depending on the attached AC adapter.
*6 The model names and specifications, etc., listed herein are subject to alteration without prior notice. Access the Shimadzu website
(https://www.shimadzu.com/an/balance/) for the latest information.
Ionizer (included as standard with the W-AD Series) specifications
Neutralization method
Alternating current corona discharge
Ion balance (*1)
± 10 V
Neutralization range
Distance (from vent): Up to approximately 400 mm
Neutralization performance (*1)
Standard time to reach 100 V or less from 1 kV charged state (distance 100 mm) Reference value: 1 second
Ozone concentration
0.06 ppm or lower (at 150 mm from the center of the vent)
Discharge electrode (Material)
Tungsten, Lifespan: Approximately 30,000 hours
Usage environment
0°C~40°C, 25%~85% RH or less (with no condensation)
Rated power supply
DC 24 V, 1.0 A
Input power supply (AC adapter)
AC 100 V, 0.58 A, 50 / 60 Hz
Contamination class (*2)
2
Overvoltage category (*2)
II
Installation environment
Restricted to indoor use
*1 Representative value measured 100 mm from center of vent with CPM (charge plate monitor) using 150×150 mm plate at 20 pF.
(At time of shipping from factory))
*2 Refer to the Shimadzu Corporation website (https://www.an.shimadzu.co.jp/balance/) for details.