background image

- 13 -

(2) One-Byte Data 

One-byte data in the text has a fixed length of one character
which is either 0 or 1.  Seven items are covered, namely, stop
status, manual status, AT status, AH status, AL/HB status
and communication mode. 

Stop (STBY), Manual (MAN), AT (ON), AH (ON),
AL/HB (ON), SB (ON), Communication (when the
remote mode is on). 

Execution (EXEC), Auto (AUTO), AT (OFF), AH
(OFF), AL/HB (OFF), SB (OFF), Communication
(when the local mode is on). 

(3) Special Data 

Special data is data sent in response to numerical data which
are outside of the ordinary range.  Their patterns are shown
in the following. 

Numerical data is +10000~+10999. 

Numerical data exceeds the indicated range.

Numerical data is over the scale. 

CT temperature exceeded the higher limit value or fell
below the lower limit value in the case of thermocouple
input.

Wire breaking detected in the case of R.T.D. input 

CT input value exceeded the higher limit value or fell
below the lower limit value.

Control output either turned on or turned off while CT
input measurement was ongoing.

Data is undeterminable due to a set value of alarm code,
P=OFF, or some other reasons. 
• Data in the case where the alarm code is 0 and D2

(AH, AL) or D3 (CT, HB) command has been
received. 

• Data in the case where the alarm code is 1~4 and D3

(CT, HB) command has been received.

• AL data in the case where the alarm code is 5~8 and

D2 (AH, AL) command has been received.

• I, D, SF or MR data in the case where P=OFF and D5

(P, I, D, SF, MR) command has been received.

7-5-3. Communication error

When an error is found in the block after receiving data from the
host, an error message is sent back.  The format of error messages
and the types of errors are described in the following.

Error Message Format 

Error Types

7-5-4. Non-response process 

When an error is found in the block upon receiving data
transmitted from the host, the next correct data block is waited for
without transmission of response data.  Therefore, the host has to
retransmit the command block in any of the following cases:

The start character was not @ (40H). 
The communication No. did not match. 
The end character was not CR (0DH).
A hardware error such as a framing error, overrun or a
parity error occurred.

7-5-5. Restrictions by commands, etc. 

(1) Restrictions by Read Command

Upon receiving the D1 read command, the status of
unequipped options (AH, AL/HB, SB) become 0 (30H). 

(2) Restrictions by Write Command 

In the case of P=OFF and Y and P output, write data of
E2 (write command of manual output value) is either of
the following two.  Otherwise, ER09 is sent back. 

In the case of P=OFF and V and I output, a lower limit
output limiter value or a higher limit output limiter
value is to be set.

If a write command is invalid due to the local mode, a set
value of keylock, AT or some other conditions, and that
command is received, ER11 should be sent back.  The
following cases also require ER11 to be sent back. 

1

0

10000

U

0

0

0

0

0

10999

U

0

0

9

9

9

more than 
11000

D

E

H

0

0

0

less than 
-3000

D

E

L

0

0

0

The (+) side

H

0

0

0

0

0

The (-) side

L

0

0

0

0

0

C

J

H

0

0

0

C

J

L

0

0

0

The higher 
limit side

The lower
limit side

+

0

0

0

0

0

The higher 
limit side

The lower
limit side

H

B

H

0

0

0

H

B

L

0

0

0

B

0

0

0

0

0

@

40H

Address No.

:

3AH

BCC

CR

0DH

Higher
position

Lower
position

Higher
position

Lower
position

E

45H

R

52H

Space

20H

Error No.

Error No.

Error Type

Description

05

06

08

09

11

12

BCC error

Command error

Data format error

Data error

Write mode error

Option error

Arithmetic value is not in accord with BCC.

Command other than prescribed one was received. 

The format is different from prescribed one.

Command of unequipped option was received. 

Write command was received when write command 
was invalid. 

Value is out of prescribed data range in the case of 
write command.

ER09

0%

+

0

0

0

0

0

100%

+

0

0

1

0

0

ER11

Condition

Received Write Command

Stop mode (STBY)

Auto mode (AUTO)
Manual mode (MAN)

E2 (Manual control output value)

E4 (Manual status)

E5 (AT status)

E2 (Manual control output value)

E5 (AT status)

E6 (Set value of AH)

E7 (Set value of AL)

E8 (Set value of HB)

E8 (Set value of HB)

E7 (Set value of AL)

EE (DF: Hysteresis) 

E5 (AT status)

EB (I: Integral time) 

EC (D: Derivative time)

ED (SF: Target value function)

EF (MR: Manual reset) 

EF (MR: Manual reset) 

ED (SF: Target value function)

Alarm code = 0

Alarm code = 1~4

Alarm code = 5~8

P=OFF

I=OFF

I=OFF

P=OFF

Summary of Contents for SR73A Series

Page 1: ...o the instrument On the terminal nameplate affixed to the case of your instrument the mark is printed This is to warn you of the risk of electrical shock which may result if the charger is touched while it is energized A means to allow the power to be turned off such as a switch or a breaker should be installed in the external power circuit to be connected to the power terminal of the instrument F...

Page 2: ... 14 8 Specifications 15 4 1 Introduction 1 1 Check before use This product has been fully checked for quality assurance prior to shipment Nevertheless you are requested to make sure that there is no error damage or shortage of delivered items by confirming the model codes and checking the external view of the product and the number of accessories Confirmation of Model Codes Check the model codes s...

Page 3: ...th 48 height 91 6 width 44 6 Dimension of current transformer CT for heater break alarm For 0 30A CTL 6 S For 0 50A CTL 12 S36 8 unit mm 2 5 Wiring 1 Wiring operation should be done according to the instruction for the terminal arrangement in section 2 6 Exercise care that no wrong connection is made 2 Crimp terminal should accommodate the M3 5 screw and should have a width of less than 7mm 3 For ...

Page 4: ...nitor LED green On selection by ON turns on AT waiting key flashes on AT execution 3 AH alarm output monitor LED red Turns on for higher limit alarm output ON 4 AL HB alarm output monitor LED red Turns on for lower alarm output ON or heater break alarm ON 5 MAN manual control output monitor LED green Flashes when control output is in manual operation 6 STBY control output stop monitor LED green Tu...

Page 5: ... decimal point in the rightmost position on the top row flashes Press the or key to change the lower limit value and press the key to register the data Upon registering the lower limit value data the decimal point in the rightmost digit in the bottom row begins to flash Change the higher limit value by pressing the or key and register it by means of the key Upon registering the higher limit value ...

Page 6: ...ualoperation 0 100 Themeasuredvalue PV isdisplayedonthetopandthecontroloutput valueonthebottom PressingtheMANkeyonthisscreenswitches autotomanualandviceversa RefertoItem4 6 1 fordetail Control execution stop selecting screen Initialvalue EXEC Settingrange EXEC STBY Execution orstop ofcontrolisselected RefertoItem4 6 2 fordetail AT auto tuning action control screen Initialvalue OFF Settingrange OFF...

Page 7: ...ockissetonlockNo 1 2or3 noselectioncanbeperformedevenifthedigitpointis flashing Digits and arenotshownincaseofnooption Digitsof and arenotdisplayed Measuring range code select digit Initialvalue Multi05 Voltage V 81 Current95fixed Selectingtwodigitsfrom 4 5 Measuringrangecodetable Control output characteristic select digit Initialvalue RA heatingcharacteristics DA coolingcharacteristics RefertoIte...

Page 8: ...the higher limit value is forced to be the lower limit value 1 Therefore the higher limit value cannot be set to be less than the lower limit value 1 ENT Note Moving to another screen while maintaining the manual mode is possible In this case it should be noted that control output is made manually The apparatus is in the manual mode when the MAN monitor LED is flashing ENT 6 1 14 Soft Start Time S...

Page 9: ...m option is added and the lower limit alarm of the alarm type select digit on the function selecting screen of the mode 2 group screen group is selected Lower limit deviation value alarm is output for measured value to be less than set value alarm set value For lower limit deviation value alarm if set value alarm set value is less than the lower limit of the measuring range the action point is low...

Page 10: ...nd 74A series two types of communication interface RS 422A and RS 485 are available Each enables a personal computer or the like to set and read various data by the use of signals which conform with EIA standards RS 422A and RS 485 are communication standards established by the Electric Industries Association EIA which are meant for electrical or mechanical matters the so called hardware but not f...

Page 11: ...ront panel turns on and the commands for read and write through communication become valid Front key operation works only to change from remote to local 7 3 2 Communication address setting screen RS 422A and RS 485 allow connection of a plurality of apparatuses but in fact communication is carried out between two apparatuses Therefore the address of each apparatus is designated by a number between...

Page 12: ... value including a code is less than six characters 0 is inserted between the code and the figure or figures to make the value six characters When a code is excluded each of the 5 characters is a figure or a decimal point Examples of numerical data 1 Command The text format of read data contains only of a command of two characters Command Contents of Read Data PV value Execution SV value 1 Control...

Page 13: ...No did not match The end character was not CR 0DH A hardware error such as a framing error overrun or a parity error occurred 7 5 5 Restrictions by commands etc 1 Restrictions by Read Command Upon receiving the D1 read command the status of unequipped options AH AL HB SB become 0 30H 2 Restrictions by Write Command In the case of P OFF and Y and P output write data of E2 write command of manual ou...

Page 14: ...xt Data 3AH BCC CR 0DH Higher position Lower position Higher position Lower position Received data Transmitted data D1 D1 STBY MAN AH AL AT SB 0 30H or 1 31H PV value SV value Output value Received data Transmitted data D2 D2 Set value of AH Set value of AL Received data Transmitted data D3 D3 CT current value Set value of HB Received data Transmitted data D4 D4 Set value of SB Received data Trans...

Page 15: ... Sampling cycle 0 5 sec PV bias range 200 unit PV filter 0 100 sec 0 without filter Control Control mode Auto tuning PID ON OFF control Proportional band P Off 0 1 999 9 FS Off setting On Off action Integral time I Off 1 6000 sec Off setting P PD action Derivative time D Off 1 3600 sec Off setting P PI action Manual reset MR 50 0 50 0 Valid when P OFF and I OFF Output limiter Lower limit limiter 0...

Page 16: ...tact type or the SSR drive voltage type Alarm action Heater amperage detected by externally attached CT except 0 5V DC input Alarm output On upon detection of heater break while control output is On Current setting range Off 0 1 50 0A Alarm action stops when Off is set or Off 1 500A when 0 5 V DC for CT input is selected Setting resolution 0 1A or 1A Amperage display 0 0 55 0A or 0 550A Display ac...

Reviews: