![shelolab SRI21D Installation & Operation Manual Download Page 30](http://html.mh-extra.com/html/shelolab/sri21d/sri21d_installation-and-operation-manual_1255384030.webp)
30 |
P a g e
OPERATION
Continued on next page
Put the unit in manual night mode (continued)
4.
Advance to the Event 1 parameter.
a.
Press the green
Advance key
3 times to reach the
parameter.
Event 1 Parameter: On
5.
Change the parameter to OFF.
a.
Press the Up Arrow once.
•
This parameter controls the chamber day-illumination
lights.
Event 1 Parameter: Off
6.
Save and advance to the Event 2 parameter.
a.
Press the Advance key once.
•
This simultaneously saves the Event 1 parameter as Off
while advancing. The small, red 2 indicator will go dark.
Event 2 Parameter: On
7.
Change the Event 2 parameter to Off.
a.
Press the Up Arrow once.
•
This parameter toggles two outputs to turn off the day
mode.
Event 2 Parameter: Off
X3