31
|
P a g e
OPERATION
P
OWER
E
XHAUST
B
LOWER
Note:
Exposure to sustained oven chamber temperatures above 80°C will damage the exhaust
blower while it is turned off. Leave the oven exhaust vent dampener closed until it is time to
turn on the blower and actively vent the oven chamber.
SHEL LAB offers an accessory forced-air power exhaust blower that can be mounted directly on the
exhaust vent cover.
Note
: Running the exhaust blower will significantly impact the oven chamber
temperature and should be used after completing a heat application.
Mounting the Exhaust Blower
x8
1.
Remove the 8 screws from the exhaust vent
cover assembly on the top, right side of the
oven.
Leave the cover assembly in place.
2.
Mount the blower on the vent cover,
positioning the exhaust output toward the back
of the oven. Ensure that the exhaust blower and
assembly screw holes are aligned.
3.
Reinstall the 8 screws to secure the exhaust
blower and vent assembly to the unit.
4.
Attach flexible ducting to the exhaust blower
to vent exhaust away from the workspace.
5.
Plug the exhaust blower into the 220 – 240
volt receptacle on the back of the oven.
Continued on next page