73
|
P a g e
UNIT SPECIFICATIONS
Temperature performance continued
Time to Temperature
: From an ambient temperature of 20°C
Model
Heat up to 80°C
Heat up to 150°C
Heat up to 220°C
SVAC1
40 Minutes
70 Minutes
120 Minutes
SVAC2
70 Minutes
120 Minutes
200 Minutes
SVAC4
70 Minutes
120 Minutes
230 Minutes
Cooldown Times
: Time to cool down to 50°C
Model
From 80°C
From 150°C
From 220°C
SVAC1
90 Minutes
122 Minutes
181 Minutes
SVAC2
110 Minutes
188 Minutes
233 Minutes
SVAC4
161 Minutes
318 Minutes
420 Minutes
P
OWER
Model
AC Voltage
Amperage
Frequency
SVAC1
110 – 120
7.0
50/60 Hz
SVAC2
110 – 120
10.0
50/60 Hz
SVAC4
110 – 120
13.0
50/60 Hz