MAINTENANCE
SDRAP Diagnostic Questions
Record the answers in the log on page 64.
Setpoint
?
What is the current temperature setpoint
?
D
isplay?
What chamber temperature is presently showing on the temperature display?
R
eference?
What temperature is the reference device presently showing for the chamber temperature?
A
mbient?
What is the current room temperature?
For best results, measure the temperature in the same
section of the room where the unit is located. Do not place your thermometer on the unit.
P
ilot Lights?
1) Is the heating active indicator on the control panel flashing or otherwise illuminating, Y/N?
2) Is the High Limit cutoff active or has it activated recently, Y/N?
Chamber Temperature in Red
Setpoint in Green
“2” indicates the
controller is calling for
power to the element
Alternating alert screens flash when the high limit heating cutoff is active.