MAINTENANCE
End of Procedure
Temperature Calibration Continued
7
Compare the reference device reading with the chamber
temperature display.
•
If the reference device and the chamber temperature
display readings are the same or the difference falls
within the range of your protocol,
the incubator is now
calibrated for temperature
.
-OR-
•
See the next step if the readings fail to match or fall
outside of your protocol range.
Reference Device
8
If the difference still falls outside the acceptable range of your
protocol, repeat steps 3 – 7 up to two more times.
Reference Device
x
9
If the temperature readings of the incubator temperature display and the reference device
still fall outside your protocol after 3 calibration attempts, contact your distributor or
customer support for assistance.