OPERATION
Note:
Each SCO10A incubator chamber CO
2
setpoint must be set independently.
S
ET THE
CO
2
S
ETPOINT
The incubator comes from the factory with the CO
2
setpoint set to Off. Set the setpoint percentage
to the gas concentration required by your study or production protocol.
Low CO
2
(LO)
When starting with no CO
2
in the chamber, the display will show “LO” until the incubator registers a
concentration greater than 0%. A few minutes of CO
2
injections may be required to build up a
sufficient gas concentration.
End of Procedure
1.
Navigate to the CO
2
Setpoint Adjustment mode
Briefly push and release either the
Up
or
Down
arrow
buttons to activate the CO
2
setpoint adjustment mode.
•
The display will briefly flash the letters “SP”,
then show the flashing, adjustable CO
2
setpoint.
Note:
The display will automatically exit the adjustment mode after 5
seconds of inactivity, with the last shown setpoint value saved.
Setpoint
Adjustment Mode
Initial Setpoint
2.
Set the CO
2
Setpoint
Use the
Up
and
Down
arrow buttons to change the
CO
2
setpoint.
Note:
To turn the CO
2
system off, set the setpoint to its
lowest setting (OFF).
New Setpoint
3.
Wait for 5 seconds after entering the Setpoint
•
The display will stop flashing, and the setpoint
is now saved in the controller.
•
The chamber will now automatically begin
injecting CO
2
or allowing the current
concentration to decay to match your setpoint.
•
The display will revert to showing the current
chamber CO
2
concentration.
CO
2
Injecting