Page 20 of 24
8.7.2.5.4 -
REV Preset
Predetermines the negative totalizer value
Default: 0000000000
Range: 1÷9999999999
Press the
key from run mode: the display will be as shown here next.
With
or
change the digit, with
moves the cursor.
Press
to confirm, the display will be as shown here next.
press the
key to exit and confirm the selection; press the
key to exit and
delete the selection.
8.7.2.6 -
Load Setting
Load factory settings.
Default: NO Range: YES - NO
Press the
key from run mode: the display will be as shown here next.
With
or
select the parameter setting.
Press
to confirm, the display will be as shown here next.
press the
key to exit and confirm the selection; press the
key to exit and
delete the selection.
8.7.3 Calibrations (Calibration)
Press the
key from run mode, then press the
key to select "Calibration" and
press the
button to enter.
8.7.3.1 -
Zero Trim
Zero flow measurement calibrate. The sensor must be full and the flow stopped.
Default: NO Range: YES - NO
Press the
key from run mode: the display will be as shown here next.
With
or
select the parameter setting.
Press
to confirm, the display will be as shown here next.
press the
key to exit and confirm the selection; press the
key to exit and
delete the selection.
8.7.3.2 -
Tube Trim
Press the
key to enter the submenu “Tube Trim” .
8.7.3.2.1 -
Full Trim
Performs a full pipe recognition self calibration.
Default: NO
Range: NO - YES
Pmag
- Setting