5-2-2. Setting “Capacity of weighing Scale”
After displaying “CAPA”, input max capacity with keys & Press
key to save & move to next
step.
If you want that set Max capa is 1,000kg, then just input “1000”.
You don’t need to consider Division value.
5-2-3. “Decimal Point” and “Digit / Division” Value
After “DIVI” displays, set Decimal point with
key.
Whenever pressing
, Decimal point will be changed. Please stop on optional position.
And set Division values with
key, Finally press
key to save.
Max. Decimal point will be 0.001, and digit can be selectable among
1, 2, 5, 10, 20, 50.
Digit and Decimal point must be fulfill the below condition.
- (Max. capacity value / division value) cannot be over than 20,000.
If this condition is not fulfilled, “Err-1” will be displayed and move back to
Capacity setting mode.