- 12 -
SI 460A DIN SIZE WEIGHING INDICATOR
5-2. Test Weight Calibration Mode
5-2-1. Calibration
Adjust weight balance between “Real weight” on the load cell(Weight Part) and
“Displayed weight of Indicator”. When you replace LOAD CELL or Indicator, you
have to Calibrate process once again.
(When you start calibration mode, TARE, HOLD & PRINT will be reset.)
Before processing calibration, please warm up the indicator during 15 min to
guarantee more preciseness.
Calibration Key Operation
CANCLE/BACK
Move to left
Move to right
Increase set
value
Save and Move
to next step
5-2-2. Start Test Weight Calibration Mode
Remove “CAL-BOLT” on the Front panel, .
and press “CAL - LOCK S/W” inside.
When “CALIBR” displays, press
key,
Then Test Weight Calibration Mode starts.
※
If you set password through “F95”, you have to input the pass word.