- 44 -
SI 460A DIN SIZE WEIGHING INDICATOR
6-1-8. Read Command Detail
Current Weight
ASCII :
STX ID(2byte) RCWT ETX
HEX :
02 30 31 52 43 57 54 03
Response
STX ID RCWT
State1(1byte) State2(1byte)
P
decimal point(1byte) +/-
(1byte) Current weight(7byte) Weight unit(2byte)
ETX
State1 : O(Over Load) , S(Steady), U(Unsteady)
State2 : N(Net weight), G(Gross weight)
Ex) Steady(S), TARE not used(N), 0.000kg
Indicator memory data
ASCII :
STX ID(2byte) RCWD ETX
HEX :
02 30 31 52 43 57 44 03
Response
STX ID RCWD P
decimal point(1byte) date(6byte) Time(6byte)
No. of weighing(6byte) +/-(1byte) Tare(7byte) +/-(1byte)
weight(7byte) weight unit(2byte)
ETX
Ex) DATE : Aug 12
th
,2014, TIME : 12:00:00, the no. of weighing : 10, TARE : 2.000kg,
current weight : 3.000kg