DIGITAL WEIGHING INDICATOR
SI 4000
14
6. Press
key to save and move to next step.
p 20.000
7
. Define the optimal Digit/Division value of weighing measurement.
Whenever you press
key, the Digit/Division value will
be changed in order of “1
2
5
10
20
50” .
Ex) Load cell CAPA : 20kg, division : 0.001
Input division “1”
d 1
8. press
key to save the Digit/Division value and move
to next step
d 1
※
Caution : (Division value /Max capacity value)
cannot over
1/20,000
.
If the division is so small compare with max capacity, Error message
“ Err 01 “
will be displayed
and move back to
“CAPA”
step again.
9. When you press
key, the indicator starts the dead process
to find “Zero” span.
dead
10. Indicator will search “DEAD weight” during 5seconds.
After find optimal “Zero” span , step is automatically
Moves to next.
Cal-1 0
※
Caution:
At this step, if there is some force or vibration on weighing scale, and unstable condition will be
continued,
“ErrorA”
will be display, and “DEAD value” will not be calculated.
Please remove all the force or vibration and process it again.
11. Span Calibration mode start.
l 20.000