- 56 -
SI 410 WEIGHING INDICATOR
7-2. Calibration Process
Display
Cause
Treatment
Eerr01
When Max capacity/digit value is over
20,000
Re-input the Max Capacity, less than
20.00
(Max Capacity / Digit)
Eerr04
Standard weight value is over than Max
Capacity
Re-input Standard weight value with
Number keys, under Max Capacity
Eerr05
Standard weight value is less than 10% of
Max Capacity
Re-input Standard weight value with
Number keys, more than 10% of Max
Capacity
Eerr06
Amp. Gain is too big
Check standard weight’s weight with set
value.
If there is difference between set value
and real weight, please re-input the value
(set value is too small)
Eerr07
Amp. Gain is too small
Check standard weight’s weight with set
value.
If there is difference between set value
and real weight, please re-input the value
(set value is too big)
Eerr08
Under “F-function” model, set value is
“N.A”
Check the correct value and re-input
err-a
When there is continuous vibration on the
weighing part,, indicator cannot process
calibration any more.
- Find vibration cause and remove
- Load cell check
- Load cell cable and connecting
condition check