- 30 - 

 

SI 410 WEIGHING INDICATOR 

5-3-4. Hidden Function   

 

How to enter Hidden function setting mode : Press 

  key 4 times and input your password. 

Default password is 

  (0410). 

Serial Number Check 

HF01 

Check your device’s serial number 

Operation time check

 

HF02 

Check how long hours it has been operated. (Power ON) Unit : 1hour 

S/W Version Check 

HF03 

Check the currently applied program version. 

H/W Version Check 

HF04 

Check the currently applied hardware version. 

DATE(Y,M,D) Check / Modification 

HF05 

Check the date or adjust when it is wrong. 

TIME(H,M,S) Check /    Modification (24Hours) 

HF06 

Check the time or adjust when it is wrong. 

Password Setting (4digit) 

HF07 

Password is required when you enter to hidden function. 
Enter the password twice. 

 

1              2              3              4              5            6              7              8              9              0 

Password combination within 

0~9.

 

Maximum Capacity Weight Check 

HF08 

Check the max capacity which is set under calibration mode. 

Division Check 

HF09 

Check the division which is set under calibration mode. 

Test weight Check 

HF10 

Check the weight of test weight which is used for your last calibration.

 

Analog value of ZERO 

HF11 

Check the analog value of ZERO. 

Analog Output Use Setting 

HF13 

●  00  4-20mA out 

 

01  0-10V out 

 
 

Summary of Contents for SI 410

Page 1: ...Digital Weighing Indicator SI 410 User Manual Ver 1 0 140825...

Page 2: ...up 13 5 1 TEST Weight Calibration Mode 13 5 2 Simulation Weight Calibration 17 5 3 F FUNCTION Setting 21 5 4 Test Mode 39 6 Interface 40 6 1 Serial Interface 40 6 2 External Input 46 6 3 Relay output...

Page 3: ...ngly used Copy Rights 1 All Right and Authority for this Manual is belonged to SEWHA CNM CO LTD 2 Any kinds of copy or distribution without permission of SEWHA CNM CO LTD will be prohibited 3 This man...

Page 4: ...age or heavy electric noise condition 4 When you connect with other devices please turn off the power of item 5 Avoid from water damage 6 For the improvement of function or performance we can change i...

Page 5: ...perating Temperature Range 10 40 14 104 Operation Humidity Range 40 85 RH Non condensing Function Calibration Mode Test Weight Calibration Mode Simulation Calibration Mode Display 6 digit 25 4mm 1inch...

Page 6: ...Port RS 422 Option2 Serial Port RS 485 Option 3 Serial Port RS 232 Option 4 ETHERNET CARD Option 5 Analog Output 0 20mA Option 6 Analog Output 0 10V Option 7 BCD OUT Option 8 BIN IN Option 9 SD Memory...

Page 7: ...ted ON OUT3 When OUT3 Relay is operated ON OUT4 When OUT4 Relay is operated ON OUT5 When OUT5 Relay is operated ON OUT6 When OUT6 Relay is operated ON OUT7 When OUT7 Relay is operated ON RUN When the...

Page 8: ...unction 1st input TARE 2nd input TARE Reset When HOLD or weight value is ZERO then this key doesn t work Number 2 Set the HOLD Function 1st input HOLD 2nd input HOLD Reset Number 3 Product number sett...

Page 9: ...weight during 5 secs Display the Sub total weight during 5 secs Display the Grand total weight during 5 secs Print the Grand total out Delete the Sub total weight Delete the Grand total weight Max acc...

Page 10: ...common Terminal O1 O2 O3 O4 O5 O6 O7 OC Relay output RELAY 1 RELAY 2 RELAY 3 RELAY 4 RELAY 5 RELAY 6 RELAY 7 RELAY COM 6 Analog output terminal Terminal 4 20mA Option 0 10V Option 7 Option serial inte...

Page 11: ...11 SI 410 WEIGHING INDICATOR 4 Installation 4 1 External Dimension Cutting Size 4 2 Installation Components SI 410 User Manual Cutting Size...

Page 12: ...tly Do not weld near the load cells Indicators or other devices Load Cell Installation 1 You can connect Max 8pcs of same capacity Load cells at once 350 2 You have to make horizontal balance on the g...

Page 13: ...ress CAL LOCK S W inside When CALIBR displays press key select WCAL and press key 5 1 3 Setting Capacity of weighing Scale If you want to set Max capacity as 1 000kg and the division is 0 1 100g then...

Page 14: ...e cannot be over 1 20 000 If this condition is not fulfilled err 1 will be displayed and move back to capacity setting mode 5 1 5 Measuring the DEAD Weight of Weighing Scale After DIVI is displayed lo...

Page 15: ...lue will not be calculated Please remove the cause of the force or vibration and process it again 5 1 6 Calculating span value If the count is over input the weight of your Test Weight and press key I...

Page 16: ...pan value during 10 20 secs After calculation span value will be displayed on the display Then press key When CALEND is displayed and calibration is completed We recommend to proceed this span value c...

Page 17: ...is calibration mode uses Load cells max capacity and Max Output Rate mV so the weight adjustment degree might be less than Test weight Calibration The guaranteed resolution of this Simulation Calibrat...

Page 18: ...load cells and each load cell s Max capacity is1 000kg Then total Max Capacity will be 4 000kg 1 000 x 4 and you have to input 4 000 5 2 3 Setting Digit Division value After CAPA displayed Check Max C...

Page 19: ...ltage mV After mV V is displayed Check the Rated output value of Load cell Refer to the load cell label or Test Report And Press key to save and move to next step When DEAD displays Press key then ind...

Page 20: ...the rated output will be same as single load cell s Because plural load cells are connected with parallel connection the sum of rated output voltage is same as single load cell s rated output Due to s...

Page 21: ...SETUP is displayed press key 1 Function No or input the function No by number key 0 9 after select Function No area by pressing key 2 Input the Set value by number key 0 9 after select Set value area...

Page 22: ...1time per 1sec 09 Fast 60times per 1sec 105 Main display setting 00 00 Weight 01 Sub total weight 02 Grand total 108 Buzzer sound External input detection 00 00 Buzzer sound 01 No Buzzer sound 109 Fu...

Page 23: ...tive within 100 of Max Capacity 211 Auto Zero function under Tare state 00 00 Disuse 01 Use 212 Tare Delay Time 00 00 Disuse 01 10 Use Unit 1sec 213 Auto tare set when weighing starts 00 00 Disuse 01...

Page 24: ...SP1 02 SP2 03 SP3 04 SP4 05 Finish 06 Zero 07 Error 230 Relay Output 5 Setting 00 00 Disuse 01 SP1 02 SP2 03 SP3 04 SP4 05 Finish 06 Zero 07 Error 231 Relay Output 6 Setting 00 00 Disuse 01 SP1 02 SP2...

Page 25: ...Stop Packer mode 10 Start Stop Packer mode 11 Print 12 Sub total Print 13 Grand total Print 14 Forced discharge 238 External Input 6 Setting 14 00 Disuse 01 Zero 02 Tare 03 Tare reset 04 Tare Tare res...

Page 26: ...der Stream Mode 00 00 Format1 19byte 01 Format 2 22byte 02 Format 3 17byte 03 Format 4 22byte 04 Format 5 15byte 306 Date transference under stream mode 00 00 Countinuously 01 Single time on every ste...

Page 27: ...us RTU 317 Check Sum under Ethernet Option Port 00 00 Disuse 01 Use 318 Data Format under Ethernet Stream Mode Ethernet Option Port 00 00 Format1 19byte 01 Format 2 22byte 02 Format 3 17byte 03 Format...

Page 28: ...PACITY Gross weight under Tare 404 BIN IN Product number change 00 00 Disuse 01 Units digit tenth digit separation 02 Units digit tenth digit No separation 405 IP setting 1 0 255 406 IP setting 2 0 25...

Page 29: ...weight Steady weight Steady weight 02 Auto At the first steady states Recent stable weight Steady weight Steady weight 03 Manual Auto At every steady states Recent finish weight Finish weight Finish w...

Page 30: ...HF05 Check the date or adjust when it is wrong TIME H M S Check Modification 24Hours HF06 Check the time or adjust when it is wrong Password Setting 4digit HF07 Password is required when you enter to...

Page 31: ...imum analog output can be adjusted key for value setting Input range 20 20 Default 0 Maximum Analog Output Setting HF15 Maximum analog output can be adjusted key for value setting Input range 20 20 De...

Page 32: ...save the date after t1 time t2 Finish Relay Output Time Function 240 Relay output Relay Condition Relay Condition OUT 1 Current weight SP1 ON Current weight SP1 OFF OUT 4 Current weight SP4 ON Current...

Page 33: ...save the date after t1 time t2 Finish Relay Output Time Function 240 Relay output Relay Condition Relay Condition OUT 1 Current weight SP1 ON Current weight SP1 OFF OUT 4 Current weight SP4 ON Curren...

Page 34: ...t date after t3 under Function 103 3 or 103 6 t3 t4 Weight Judge Time Function 243 Relay output Relay Condition Relay Condition OUT 1 Near zero Steady weight SP1 ON OUT 4 SP3 Steady weight SP4 ON OUT...

Page 35: ...elay Condition Relay Condition OUT 1 Near zero Current weight SP1 ON OUT 5 SP4 Current weight ON OUT 2 SP1 Current weight SP2 ON OUT 6 Within near zero range Function 201 ON OUT 3 SP2 Current weight S...

Page 36: ...Condition OUT 1 Start On ON Current weight SP1 OFF OUT 4 Start On ON Current weight SP4 OFF OUT 2 Start On ON Current weight OFF OUT 5 After SP4 After t1 set time During t2 set time ON OUT 3 Start On...

Page 37: ...y Output Time Function 240 Relay output Relay Condition Relay Condition OUT 1 Current weight SP1 ON Current weight SP1 OFF OUT 4 Current weight SP1 SP2 SP3 SP4 ON Current weight SP1 SP2 SP3 SP4 OFF OU...

Page 38: ...ate after t1 time t2 Finish Relay Output Time Function 240 Relay output Relay Condition Relay Condition OUT 1 Start input ON Current weight SP1 OFF OUT 4 SP1 SP2 SP3 Current weight SP1 SP2 SP3 SP4 ON...

Page 39: ...ted devices Test Mode Analog Value Press key 4times Analog Variation Value Press key 4times Key Press key 4times Display Press key 4times External Input Press key 4times Relay Output Press key 4times...

Page 40: ...DY GS when setting TARE US UNSTEADY 2 Data Format 2 ID Number Data Transference Refer F function 101 305 01 22byte ID Number Function 101 Header1 Header2 OL OVER LOAD NT NET WEIGHT Tare is not set ST...

Page 41: ...unction 101 Header1 Header2 OL OVER LOAD NT NET WEIGHT Tare is not set ST STEADY GS when setting TARE US UNSTEADY Bit 0 Bit 1 Bit 2 Bit 3 Bit 4 Bit 5 Bit 6 Bit 7 ZERO TARE Gross Weight Print HOLD 1 ST...

Page 42: ...T ETX 22 byte 22 byte Current data STX ID RCWD ETX 44 byte 48 byte Sub total date STX ID RSUB ETX 29 byte 30 byte Sub total weighing times STX ID RSNO ETX 14 byte 14 byte Grand total data STX ID RGRD...

Page 43: ...WGPR ETX 8 byte Delete Sub total STX ID WSTC ETX 8 byte Delete Grand total STX ID WGTC ETX 8 byte Run STX ID WSTR ETX 8 byte Stop STX ID WSTP ETX 8 byte Time setting STX ID WTIM TIME HHMMSS ETX 14 by...

Page 44: ...Multiple Registers CRC Check Method is CRC 16 Add ress Leng th Featur e Description 1 2 RO Capacity 3 2 RO None 0x00 5 2 RO Analog Value 7 2 RO Span Value 9 1 RO Division 10 1 RO Decimal point 11 2 RO...

Page 45: ...ro Tare OUT1 OUT2 OUT3 OUT4 Hold 3 Error register Address 21 Length 2 0 1 2 Loadcell Error Over Load Set point Error 4 Key register Address 441 Length 1 0 1 2 3 4 5 6 7 8 9 10 11 Start Stop Zero Tare...

Page 46: ...y output 7pcs relay out A dry Each relay output s function is selectable from function 226 232 6 3 1 Specification Contact Ratings VDC Contact Ratings VAC 24V 3A 250V 3A 6 3 2 Relay output circuit com...

Page 47: ...loop is suitable for middle distance transmission because stronger than RS 232C against electric noise About 100M Maximum communication speed is 9 600 6 4 1 Current loop circuit composition 6 4 2 Con...

Page 48: ...ogue output value is adjusted with DIGITAL MULTI METER 3 How to adjust analog output value Step1 Connect Digital multi meter to the Indicator A out terminal Step2 Enter F function HF14 Minimum Analog...

Page 49: ...t analog output value is 0V weight zero 10V Full capacity 2 The analogue output value is adjusted with DIGITAL MULTI METER 3 How to adjust analog output value Step1 Connect Digital multi meter to the...

Page 50: ...0 10V 2 HF13 Analog output setting should be changed also 6 8 Print Interface It can be connected with all kinds of Serial interface printer but the printing format is already programmed and fixed wit...

Page 51: ...IN IN card Changing Product number 6 9 1 BIN IN card circuit composition 6 9 2 BIN IN card connection PIN No 1 14 2 15 3 16 4 17 5 Role IN1 IN2 IN3 IN4 IN5 IN6 IN7 IN8 IN COM Function 404 00 1 2 4 8 1...

Page 52: ...4X10000 28 2X1 2 8X10000 11 4X1 20 1X100000 29 8X1 3 2X100000 12 1X10 21 4X100000 30 2X10 4 8X100000 13 4X10 22 COM 32 8X10 5 Net weight HIGH 31 1X100 23 Total weight LOW 31 2X100 6 GND 1 14 4X100 24...

Page 53: ...ard Weighing data will be saved to SD memory card depends on your function 103 6 12 1 Saving format File name YYMMDD CSV Ex 140728 CSV DATE TIME ID PART SERIAL WEIGHT UNIT 2014 07 18 12 18 04 1 50 22...

Page 54: ...SD card to PC and select FORMAT from PC system folder Select FAT32 from file system 6 13 Option card combination Maximum 2EA of option card can be instlled Below combination is available SERIAL 232 SE...

Page 55: ...Isolate Resistance is more than 100M Weight Value is increased regular rate but not return to Zero 1 Load cell Error 2 Load cell connection Error 1 Check Load cell connection 2 Measure Load cell Resis...

Page 56: ...ty Eerr06 Amp Gain is too big Check standard weight s weight with set value If there is difference between set value and real weight please re input the value set value is too small Eerr07 Amp Gain is...

Page 57: ...D board error 3 Try to connect the indicator s A D with the other indicator 4 Check the power and connection of terminal 1 Power is ON when some materials are on weighing part Under Normal Mode if the...

Page 58: ...Modification or Repair without Sewhacnm s permission Any kinds of Mal function Defection or External damage caused by operator Any kinds of Mal function Defection caused by using spare part from Non...

Reviews: