6
Operation
Turning the MP10 Display On -
The MP10 is powered on auto-
maticaly by opening the lid. The MP10 displays an opening screen for 5
seconds, after which it displays the default MicroPurge screen. At this
point the MP10 is in MicroPurge mode (MP) but not cycling the pump.
This initial state allows the user to adjust time and throttle settings
before the pump starts to operate. Pressing the Cycle key begins pump
cycling. Times and modes may be adjusted while the pump is cycling
or before. Pressing the Cycle key a second time will stop pump cycling.
Note: all user-entered time settings are lost when the MP10 is turned off.
Also, the MP10 automatically powers down when the lid is closed, so
make sure the MP10 is stored with its lid closed.
Opening Display -
The opening display is shown for 5 seconds and
displays the controller name, the version number and the battery voltage,
as shown in Figure 1. Figure 1 shows that the battery is GOOD, that the
battery voltage is 4.20 volts and that the software version in the cont-
roller is 1.0. Battery voltage must be greater than 3.6 volts for the unit
to operate. If the unit fails to cycle replace the 3-AA batteries with fresh
cells. The opening screen is displayed for 5 seconds, if you wish to by
pass the opening screen, hitting any key, such as the CPM key will bring
you to the default MP display.
Figure 1 Opening Screen
OPERATION
MicroPurge Basics Controller
MP Micro Purge
ID ID Time Set
MN Manual Time Set
LVL LevelShutoff
CYCLE
I I
MODE
CPM/Value
Flow/Value
Flow/Value
Hold/Sample
/Cycle
Start/Stop
ID/MN/MP/
Battery
Mode
CPM
ID
Time
Refill
Discharge
QED
MP10
V1.0
BAT
GOOD
4.20
D
D
E
E